Canary MicroMux Скачать руководство пользователя страница 9

 

Section 2 – MultiMux Operation and Installation 

2.8 Enclosure Installation 

 
The MicroMux can be attached to any surface with 4 mounting 6-32 screws. 
 
The placement of the mounting holes is depicted in the illustration below. 

H

1

L

1

4

H

1

4

L

1

4

H

2

V

W

 

G

a

g

e

4

L

2

CHANNEL 

4

MicroMux4

3

H

1

3

L

1

3

H

2

V

W

 

G

a

g

e

3

L

2

CHANNEL 

3

2

H

1

2

L

1

2

H

2

V

W

 

G

a

g

e

2

L

2

CHANNEL 

2

1

H

1

1

L

1

1

H

2

V

W

 

G

a

g

e

CHANNEL 

1

S

1

2

V

G

C

L

K

V

W

 

G

a

g

e

E

N

COMMON 

H

2

L

2

www.canarysystems.com

Made in the U.S.A.

CURRENT LIMIT: 2A 30VDC MAX

S

S

H

IE

L

D

  

1

L

2

6 inches

2 inches

6 inches = 15.2cm
2 inches = 5.1cm

Note: The Micromux can be installed 
upside down, vertically, or horizontally.

 

2.9 Input Protection 

 
The MicroMux is equipped with lightning protection components on the control signals. As a result, care 
must be exercised in the installation to maximize their effectiveness. Specifically, an effective earth 
ground must be attached to the MicroMux shield (S) for proper protection. Most Vibrating wire instruments 
also have lightning protection components. 
 
The Enable and Clock inputs can be connected to 12 volts directly without damaging the MicroMux. The 
connections to the datalogger should be disconnected and insulated from contacting other electrical 
connections. The 12V tolerant inputs allow the MicroMux to be field tested without the need for a 
datalogger connection. The enable can be wired to the 12V input while the clock input can be touched to 
12V. This action will cause the MicroMux to step channels for each touch to 12 volts. 
 
 
 

Содержание MicroMux

Страница 1: ...ans whether directly or indirectly without the written consent of Canary Systems This document is subject to change without notice and Canary Systems assumes no responsibility for errors omissions or...

Страница 2: ...ration Details 4 2 2 Datalogger Connection 5 2 3 Instrument Connection 5 2 4 MultiLogger Software Configuration 5 2 5 CR10X Program Example 6 2 6 CR1000 Program Example 7 2 7 Using the MicroMux with t...

Страница 3: ...irtually infinite channel isolation Warranty is applicable for 2 years from date of shipment Warranty does not cover failure by improper installation misuse or by nature including but not limited to f...

Страница 4: ...nnels would be 4 and 8 respectively The channel switching mode is selected by configuring the DIP switch mounted on the MultiMux board The table shown below describes the 2 possible configurations DIP...

Страница 5: ...del on the Configure Multiplexers form Before the individual channels may be edited you must select a Gage Type Select either 4 Channels default 8 Channels to match the DIP switch settings of the Micr...

Страница 6: ...gh 3 Excitation with Delay P22 50ms delay after enabling the MicroMux 1 1 Ex Channel 2 0 Delay W Ex units 0 01 sec 3 5 Delay After Ex units 0 01 sec 4 0 mV Excitation 4 Beginning of Loop P87 1 0 Delay...

Страница 7: ...iplexer PortSet 1 0 2 7 Using the MicroMux with the VW Comm Module The VW Comm has built in instructions to directly control the MicroMux MultiMux or MiniMux There are specific gage types in Multilogg...

Страница 8: ...0 ScratchLoc 4 FormatFloat Channel 2g Send Set Digital Outputs command Switch to Gage 1 SerialOut Com1 0MM0 ScratchLoc 4 00045 5 50 Send Take Reading command SerialOut Com1 0M 00045 5 50 SerialFlush c...

Страница 9: ...e MicroMux is equipped with lightning protection components on the control signals As a result care must be exercised in the installation to maximize their effectiveness Specifically an effective eart...

Страница 10: ...ble and Clock lines connected to the control ports Attach Enable Clock lines No Do the software settings for the Enable and Clock lines match the connections Match software and digital I O connections...

Отзывы: