Ensure the cooling water is flowing through the bath before pressing standby to switch on the 
control unit.  Press the standby key and the LED display will illuminate. The actual 
temperature of the tissue bath is displayed.   The unit is now operational and power is being 
supplied to the tissue bath cooling head.  Pressing either the ‘+’ or ‘-‘ keys once will display 
the set temperature.  Continuing to press the ‘+’ or ‘-‘ keys will change the set temperature.  
The displayed temperature will revert back to the actual bath temperature a few seconds after 
the last key press. 
 

Standby 

 

 
If the cooling bath is not connected or has a fault the unit will display ‘Err’.  In this situation, 
no power is applied to the cooling bath.   
 
The temperature feedback sensor is mounted in the floor of the cooled bath.  The temperature 
at this point will obviously be different from the temperature at the specimen holder. Over 
time this temperature difference will settle to between 1ºC and 2ºC above that of the 
measured temperature (depending on conditions). 
For the users convenience the set temperature and actual temperature displayed are offset by 
1ºC.i.e. A displayed plate temperature of 5ºC will actually be 4ºC. 
 
 

4. Cleaning & Maintenance 

 
The 7600 Control unit contains no user-serviceable parts and requires no maintenance. 
The standard bath assembly requires only cleaning after use. 

It must not be sterilised by 

autoclaving methods nor must it be immersed in water. 

  
The autoclavable specimen bath is manufactured from a special grade of Nickel-Aluminium 
Bronze, a material with excellent corrosion resistance properties. The cooling assembly must 
not be autoclaved therefore the bath unit must be separated from it before autoclaving.  The 
bath is secured to the cooling assembly with 4 stainless steel screws M3 x 35 long. When 
fitting the bath to the cooling assembly, the underside of the bath should be smeared with 
heat-conducting compound (Campden Instruments model 7600-HTP). The bath should be 
then ‘wrung’ onto the cooling head to ensure a good heat transfer path and the screws fitted 

 

6

Содержание 7600 Series

Страница 1: ...Campden Instruments UK PO Box 8148 Loughborough Leicestershire LE12 7TJ Tel 44 01509 817700 Fax 44 01509 81770 Campden Instruments USA at Lafayette Instrument Co PO Box 5729 Lafayette Indiana 47903 T...

Страница 2: ...en Instruments Limited No part of this manual may be reproduced in any form or by any means including photocopying or storing in any medium by electronic means without the written permission of the co...

Страница 3: ...l be able to advise on this matter 1 1 Overview The 7600 Tissue Bath Cooler Unit is intended for use with the Campden Instruments Range of vibrating microtomes Unfixed brain slices sectioned at 4o C g...

Страница 4: ...rature to within 0 5 C of the temperature set point There will however be a small variation in temperature vertically through the bath Experience will show the best temperature to be set for any given...

Страница 5: ...t moulding See Figure 2 On Off switch Mains socket Voltage selector fuse drawer Figure 2 Mains inlet voltage selector example shown is set to 110 120V The inlet moulding accepts a standard IEC socket...

Страница 6: ...tle to between 1 C and 2 C above that of the measured temperature depending on conditions For the users convenience the set temperature and actual temperature displayed are offset by 1 C i e A display...

Страница 7: ...of 19000ppm A c s f has a much higher concentration This concentration will increase due to evaporation and the passive oxide layer will break down For this reason it is essential that the bath and h...