Table 12-1: CPIStatus activity
Response
Meaning
Active
The module is connected to the CPI bus and is making measurements according
to the data logger program.
Offline
The module was present after startup but is no longer responding.
Unused
The module is or was connected and powered but is not included in the data
logger program.
Wait Config The module has not yet responded to a data logger’s attempts to configure it.
Config Fail
The module could not be configured. A configuration error message is appended
to this response.
12.2 Status LED
The Status LED indicates the current operation of the TEMP 120, as shown in the following table..
Table 12-2: COMM Status LED Activity
Green
Flashing
Module has been configured by data logger and is receiving sync
signals
Orange
Flashing
Device has not been configured by data logger
Red
Single-Flash Device has been configured but is not receiving sync signals
Double-
Flash
Device has encountered a scan timeout
Solid
Error
12.3 CPI port LEDs
Green and Orange LEDs on the CPI bus indicate the current operation of the CPI bus as shown in
the following table.
Table 12-3: CPI port LED activity
Green
Flashing
Frame sync
Orange
Flash
CPI activity
Solid
CPI bus fault