5. Adjusting screws are located on the bottom adjacent to the large center drain hole. Adjust
both screws the same number of turns. Rotation clockwise increases the number of tips per
16 oz. of water; counter clockwise rotation decreases the number of tips per 16 oz. of water.
One half turn of both screws causes a 2% to 3% change.
6. Check and re-level the rain gage.
Factory calibration:
If factory calibration is required, contact Campbell Scientific to obtain an RMA (see
at
back of manual).
TE525-Series Tipping Bucket Rain Gage