TB4 and TB4MM Tipping Bucket Rain Gages 

5.

 

Select any other sensors you have, then finish the remaining 

Short Cut

 

steps to complete the program. The remaining steps are outlined in 

Short 

Cut Help

, which is accessed by clicking on 

Help

 | 

Contents

 | 

Programming Steps

6.

 

If 

LoggerNet

PC400,

 

RTDAQ

, or 

PC200W

 is running on your PC, and the 

PC to datalogger connection is active, you can click 

Finish

 in 

Short Cut

 

and you will be prompted to send the program just created to the 
datalogger. 

7.

 

If the sensor is connected to the datalogger, as shown in the wiring 
diagram in step 4, check the output of the sensor in the datalogger support 
software data display to make sure it is making reasonable measurements. 

5.  Overview 

The TB4 and TB4MM tipping bucket rain gages are manufactured by 
Hydrological Services Pty. Ltd. (Model TB4) and modified for use with 
Campbell Scientific dataloggers. 

These rain gages catch rainfall in the 200 mm (7.87 in) diameter collection 
funnel. When a full bucket of rainfall is collected, the tipping bucket assembly 
tips and activates a reed switch. The switch closure is recorded by the 
datalogger pulse channel. When the bucket tips, the water drains out the 
screened fittings in the base of the gage.  

The rain gages are ideal for locations where intense rainfall events may occur. 
They include a siphoning mechanism that allows the rain to flow at a steady 
rate regardless of rainfall intensity. The siphon reduces typical rain bucket 
errors and produces accurate measurements for up to 50 cm per hour. 

The 260-953 Alter-Type Wind Screen can be used with the TB4 and TB4MM 
to minimize the effects of strong winds. Siting information and the installation 
procedure for this wind screen is provided in our 

260-953 manual

. 

6.  Specifications 

Features: 

 

More accurate measurement of high-intensity precipitation 

 

High precision—tips at 0.01-in. increments 

 

Compatible with the CWS900-series interfaces, allowing it to be used 
in a wireless sensor network 

 

Compatible with Campbell Scientific CRBasic Dataloggers: 
CR200(X) series, CR300 series, CR6 series, CR800 series, CR1000, 
CR3000, CR5000, and CR9000(X) 

Funnel

200 mm (7.87 in) 

Drain Fittings:

 

Accept 12 mm ID tubing 

Measurement Range:

 

0 to 500 mm/hr  

 

 

(0 to 19.7 in/hr) 

Accuracy:

 

Better than 

+

2% @ 500 mm/hr (19.7 in/hr) 

Содержание TB4

Страница 1: ...INSTRUCTION MANUAL TB4 and TB4MM Rain Gages Revision 4 16 C o p y r i g h t 1 9 9 5 2 0 1 6 C a m p b e l l S c i e n t i f i c I n c...

Страница 2: ......

Страница 3: ...ucts to CSI CSI will return such Products by surface carrier prepaid within the continental United States of America To all other locations CSI will return such Products best way CIP port of entry per...

Страница 4: ...g container Campbell Scientific s shipping address is CAMPBELL SCIENTIFIC INC RMA _____ 815 West 1800 North Logan Utah 84321 1784 For all returns the customer must fill out a Statement of Product Clea...

Страница 5: ...beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit cl...

Страница 6: ......

Страница 7: ...tion 6 7 3 Siting 7 7 4 Mounting 7 7 5 Leveling 8 8 Troubleshooting and Maintenance 9 8 1 Troubleshooting 9 8 2 Maintenance 10 8 2 1 Dismantling for Cleaning 10 8 2 2 Reassembling the TB4 13 8 3 Calib...

Страница 8: ...of the TB4 9 8 1 Components of TB4 Base 10 8 2 Dismantling the Filter Siphon Assembly 11 8 3 Filter Siphon Assembly 12 8 4 Reassembling the TB4 13 Table 7 1 Wire Color Wire Function Datalogger Connec...

Страница 9: ...e precision instruments Please handle them with care The black outer jacket of the cable is Santoprene rubber This compound was chosen for its resistance to temperature extremes moisture and UV degrad...

Страница 10: ...datalogger wiring terminals Short Cut is available as a download on www campbellsci com and the ResourceDVD It is included in installations of LoggerNet PC200W PC400 or RTDAQ Use the following proced...

Страница 11: ...Rain box and selecting one of the other options The default rainfall per tip value of 0 01 inches should be used if the sensor is a TB4 purchased at Campbell Scientific U S office Select 0 2 mm if the...

Страница 12: ...tips and activates a reed switch The switch closure is recorded by the datalogger pulse channel When the bucket tips the water drains out the screened fittings in the base of the gage The rain gages...

Страница 13: ...ial The 260 953 Alter Type Wind Screen s siting information and installation procedure are provided in our 260 953 manual 7 1 Wiring Connections to Campbell Scientific dataloggers are given in TABLE 7...

Страница 14: ...fter they are imported and edited in CRBasic Editor A Short Cut tutorial is available in Section 4 QuickStart p 2 If you wish to import Short Cut code into CRBasic Editor to create or add to a customi...

Страница 15: ...surface Each pad includes a hole that fits a 3 8 inch M8 bolt The three holes form a 234 mm 9 21 in diameter bolt circle Campbell Scientific offers the CM240 mounting bracket for installing and leveli...

Страница 16: ...emove the housing assembly from the base by loosening the three housing screws and lifting the housing upward 2 Adjust the three leveling screws on the CM240 bracket to level the gage FIGURE 7 4 A bul...

Страница 17: ...tion of Hazardous Material and Decontamination form Refer to the Assistance page at the beginning of this manual for more information 8 1 Troubleshooting Symptom No Precipitation 1 Check that the sens...

Страница 18: ...ach bucket tip 8 2 1 Dismantling for Cleaning The following items should be checked regularly for cleanliness Filter FIGURE 8 3 Siphon FIGURE 8 3 Tipping bucket assembly FIGURE 8 1 Top surface of adju...

Страница 19: ...pulling the siphon see FIGURE 8 2 Do not twist the filter siphon assembly while pushing and pulling FIGURE 8 2 Dismantling the Filter Siphon Assembly CAUTION Do not twist while pushing and pulling Pu...

Страница 20: ...y doing the following FIGURE 8 3 a Unscrew nut b Lightly press stem down on surface until stem pops out of siphon body c Remove stem from siphon body d Unscrew cap e Clean all items FIGURE 8 3 Filter...

Страница 21: ...assembly while putting it back into place FIGURE 8 4 Reassembling the TB4 5 Place the housing assembly back onto the base and tighten the three screws that secure the housing onto the base 8 3 Calibr...

Страница 22: ...314 cc 19 16 in3 of water over a 15 minute period into the collection funnel This volume of water is equal to 39 inches of rainfall 10 mm 4 After the water has passed through the rain gage the tipping...

Страница 23: ...ram and exit Short Cut Make note of the file name used when saving the Short Cut program 2 Open CRBasic Editor 3 Click File Open Assuming the default paths were used when Short Cut was installed navig...

Страница 24: ......

Страница 25: ...ain Gage measurement Rain_mm PulseCount Rain_mm 1 1 2 0 0 254 0 Call Data Tables and Store Data CallTable TB4_mm NextScan EndProg B 1 2 CR200 X Series Pulse Channel Example CRBasic Example B 2 CR200 X...

Страница 26: ...0 Min 0 Totalize 1 Rain_mm FP2 0 EndTable Main Program BeginProg Scan 1 Sec 1 0 PulseCount Rain_mm 1 18 2 0 254 0 Black wire connected to C8 CallTable Rain NextScan EndProg B 2 2 CR200 X Series Contro...

Страница 27: ......

Страница 28: ...ua Apinag s nbr 2018 Perdizes CEP 01258 00 S o Paulo SP BRASIL www campbellsci com br vendas campbellsci com br Campbell Scientific Canada Corp 14532 131 Avenue NW Edmonton AB T5L 4X4 CANADA www campb...

Отзывы:

Похожие инструкции для TB4