Campbell ST21 Скачать руководство пользователя страница 11

User Manual 

1.5 One 

Argos 

ID Number 

The ArgosDataRepeat() function is most useful when only one Argos ID number 
is used.  It does not matter how many data buffers use the same ID.  
ArgosDataRepeat() allows the selection of which data buffers will be transmitted, 
and at what repetition period.  When ArgosDataRepeat() is used to start data 
transmission, the ST-21 will wait one repetition period between each transmission.  
One buffer is transmitted each transmission, regardless of how many buffers have 
been enabled.  The maximum number of transmissions that can be scheduled with 
ArgosDataRepeat() is 255. 

2. Datalogger 

Programming 

There are 5 programming instructions available for use with the ST-21 Argos 
PTT: ArgosSetup, ArgosData, ArgosDataRepeat, ArgosTransmit, and ArgosError.  
The use of ArgosSetup is optional.  ArgosData and ArgosDataRepeat are used to 
enable transmission of data.  More advanced applications can use ArgosTransmit.  
If the setup parameters of the PTT need to be changed, use ArgosSetup first.  The 
setups are maintained in memory until the PTT resets.  Once the setup parameters 
are set, use ArgosData followed by ArgosAutoRepeat.  Data in a buffer is saved 
until the PTT resets or is overwritten by another occurrence of ArgosData.  If 
either the ArgosDataRepeat or ArgosTransmit instruction tries to use a data buffer 
that has not been setup, the instruction will fail.  

2.1 Argos 

Setup 

ArgosSetup (Result Code, Buffer Number, Decimal ID, Hexadecimal ID, 
Frequency)  

2.1.1 Result 

Code 

Result Code used to determine if command executed successfully.  The Result 
Code is set true on success, and false on failure. 

2.1.2 Buffer 

Number 

Buffer Number is the ST-21 buffer to setup.  Valid entries are 0 through 6. 

2.1.3  Decimal ID Number 

The Decimal ID number, and Hexadecimal ID are assigned to the Buffer Number.  
Entries must be valid ID numbers or ArgosSetup will return False.  The decimal 
ID is not a different representation of the same number as the Hexadecimal ID.  
The two ID numbers are different. 

2.1.4 Frequency 

Frequency to be assigned to the Buffer Number.  Frequency is in Hertz.  Valid 
entries: 401630000 to 401656000 in steps of 2000 Hz and 401676000, 
401678000, and 401680000 Hz. 

2.2 ArgosData 

ArgosData (ResultCode, ST21Buffer, DataTable, NumRecords, DataFormat) 

Содержание ST21

Страница 1: ...ST 21 Argos Satellite Transmitter User Manual Issued 10 12 12 Copyright 2006 2012 Campbell Scientific Inc Copied under licence by Campbell Scientific Ltd CSL 884...

Страница 2: ......

Страница 3: ...e repair is under guarantee or not Please state the faults as clearly as possible and if the product is out of the guarantee period it should be accompanied by a purchase order Quotations for repairs...

Страница 4: ......

Страница 5: ...l cell phones and aerials may also not be applicable according to your locality Some brackets shields and enclosure options including wiring are not sold as standard items in the European market in so...

Страница 6: ......

Страница 7: ...Number 3 2 1 3 Decimal ID Number 3 2 1 4 Frequency 3 2 2 ArgosData 3 2 2 1 General Description 4 2 2 2 ResultCode 4 2 2 3 ST21Buffer 4 2 2 4 DataTable 4 2 2 5 NumRecords 4 2 2 6 DataFormat 4 2 3 Argos...

Страница 8: ...ansmitter 1 1 2 The ST 21 s 9 pin connector attaches to the datalogger via the SC12 cable 2 3 1 The ST 21 s BNC connector attaches to the antenna via the antenna cable 6 3 2 The 12022 antenna 7 Tables...

Страница 9: ...Figure 1 1 ST 21 Argos Satellite Transmitter 1 1 Specifications Power supply Nominal 12 volts DC 10 volts DC minimum 16 volts DC maximum Current drain at 12 volts 1 12 mA quiescent 375 mA during trans...

Страница 10: ...ng 20 bit ID numbers 32 bytes of data can be stored in the buffer The data buffers are numbered 0 through 7 with 7 being reserved for other uses Data buffer setup can be done by the factory or under d...

Страница 11: ...in memory until the PTT resets Once the setup parameters are set use ArgosData followed by ArgosAutoRepeat Data in a buffer is saved until the PTT resets or is overwritten by another occurrence of Ar...

Страница 12: ...ds NumRecords is the number of data records that will be copied from the data table to the Argos PTT Care should be taken to make sure the total amount of data does not exceed what the buffer can hold...

Страница 13: ...smit instruction allows the datalogger to control the timing of each transmission If a buffer is set to transmit but the buffer has not been setup with valid parameters ArgosDataRepeat will fail 2 4 A...

Страница 14: ...a Connector and Antenna The ST 21 s BNC connector is for attaching the antenna see Figure 3 1 Figure 3 1 The ST 21 s BNC connector attaches to the antenna via the antenna cable The connector s RF outp...

Страница 15: ...lite passes over the poles 14 times a day providing 28 satellite passes per day Coverage decreases with latitude At the equator the PTT will see a satellite six or seven times a day The duration of th...

Страница 16: ...1 http www cls fr html argos welcome_en html 5 Antenna Placement The position of polar orbiting satellites relative to the PTT will change during the satellite pass and with each satellite pass The an...

Страница 17: ......

Страница 18: ...rta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Centro Caribe S A CSCC 300N Cementerio Edificio Breller Santo Domingo Heredia 40305 COSTA RICA www campbellsci cc i...

Отзывы: