8 SR50A Firmware Updates
The firmware on the SR50A can be updated by using a
terminal program such as Hyperterminal. If your SR50A is
configured for RS-232 communications, the SR50A can be
programmed from the I/O cable. The SR50A also contains an
internal female DB-9 connector.
Firmware updates should only be performed at an appropriate
workstation with static control procedures in place. Failure to
follow the procedures may cause damage to the sensor.
x
Ensure that the SR50A is completely disconnected.
x
Open the SR50A as follows:
o
Remove the 6 Philips head screws that are located on the
bottom side of the transducer housing.
Note: The screws used on the SR50A have changed from the
slotted type to Philips.
o
The transducer housing will separate from the main
housing.
o
Ensure that the screws and o-rings are kept and set aside.
o
Disconnect the transducer housing assembly from the
main body assembly by pressing the tab on the
connector and separating the connector.
o
Remove the 2 Philips screws from the bottom disk
assembly.
o
The bottom disk assembly should now separate from the
main housing.
o
The circuit board will still remain connected to the
housing via the signal wires.
o
Place the bottom disk/ circuit assembly on the work
bench with the circuit board and DB-9 connector
facing up. Ensure that no part of the PCB is in
contact with the lid or other conductive objects.
x
Move the jumper on the Run/Program Header from the
Run position to the Program position.
x
Connect to the DB9 connector on the SR50A