Section 2.  Storing Data from Current Dataloggers

2-2

previous output, and updates the Storage Module Pointer to the DSP
location.  This transfer occurs in binary at 9600 baud.

Under optimum conditions, the datalogger outputs 480 low resolution data
points per second.  For example, a full 128K CR10X takes over two minutes to
complete the data transfer.  If the execution interval is less than two minutes,
the data transfer will be interrupted and will take longer to complete.

Be sure to complete the data transfer before making any
changes in the datalogger program.  Changing a program
and then compiling it will advance the datalogger’s Storage
Module Pointer to the position of the DSP.  If this is done
before connecting the Storage Module to allow the
datalogger to transfer the data, the old data will not be
automatically transferred to the Storage Module.  Also, the
Instruction 96 data transfer will not take place while the
datalogger is in communication with another device (e.g., a
computer or an active CR10KD).

To be certain that the Storage Module has been connected during an execution
of Instruction 96, you can observe the green “ WRITE”  LED of the SM which
indicates when data is being stored into the module.  The user should connect
the SM to the datalogger and wait for the green LED to light.  When the LED
goes off
, data transfer is complete and the SM can be disconnected.

To avoid data loss, data must be retrieved before the datalogger’s ring memory
wraps around and begins to write over the oldest data.

2.1.1  Multiple Storage Modules with the CR500/510, CR10/10X,

or CR23X

These dataloggers use synchronous addressing to communicate with multiple
peripherals.  The CR23X must have OS1.7 or later to work with multiple
modules.  Up to four SM4M/SM16M Storage Modules can reliably be
connected to a single datalogger.  Each is assigned an address, 1 to 8.
Addresses are assigned in the *9 Mode (command 10), using SMS or in the
Telecommunications Command State (the L command).  These are discussed
in the datalogger manuals and Appendix B respectively.  The default Storage
Module address after Reset (as shipped from the factory) is #1.  This will be
sufficient for most users.

While up to eight Storage Modules can be addressed,
communication may be unreliable if more than four SM4M
or SM16M modules are connected to a datalogger at a
time.  Contact Campbell Scientific if you need to connect
more than four modules.

At a given time, only one Storage Module will accept commands directed to
SM #1.  If there is no Storage Module with address #1, the Storage Module
with the lowest numbered address will accept the commands sent to SM #1.

CAUTION

CAUTION

Содержание SM16M

Страница 1: ...INSTRUCTION MANUAL SM4M SM16M Storage Module Revision 3 01 C o p y r i g h t c 2 0 0 0 2 0 0 1 C a m p b e l l S c i e n t i f i c I n c...

Страница 2: ...ther warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential...

Страница 3: ...ation 1 7 1 5 2 Programming the Datalogger to Send Data to the Storage Module 1 7 1 5 3 Manual Dump 1 7 1 5 4 Data Retrieval 1 7 Section 2 Storing Data from Current Dataloggers 2 1 2 1 Instruction 96...

Страница 4: ...A Interface 5 1 5 1 2 Via Datalogger 5 2 5 1 3 Via Remote Datalogger and Modem 5 3 5 2 SMS 5 3 5 3 Data Retrieval 5 3 5 3 1 Direct Connection 5 3 5 3 2 Connection Via CR500 510 CR10 10X or CR23X 5 4...

Страница 5: ...ns Command State 1 5 5 1 Storage Module Connected to a Portable PC via SC532 A Interface 5 1 5 2 Communication Via the CR500 510 CR10 10X or CR23X 5 2 5 3 Communication Via the CR23X 5 2 5 4 Typical S...

Страница 6: ...SM4M 16M Flash Storage Module Table of Contents iv This is a blank page...

Страница 7: ...iew should be read by all users as should Section 5 Data Retrieval Users with the CR500 510 CR10 10X CR23X or with the 21X or CR7 with recent operating systems i e current dataloggers should read Sect...

Страница 8: ...ndby State Less than 200 A Peak current flash erase 60 mA Maintenance There are no user serviceable parts inside the Storage Module L o g a n U T S M 4 M S N S O L I D S T A T E S T O R A G E M O D U...

Страница 9: ...s Typically the flash memory can be erased and rewritten a minimum of 100 000 times 1 4 Function The SM4M and SM16M Storage Modules supplement the storage capacity in a datalogger The Storage Modules...

Страница 10: ...ble Method In the Printer Enable Method see Figure 1 3 the Storage Module stores all data received on pin 9 while the printer enable line pin 6 is high The Storage Module automatically detects the bau...

Страница 11: ...ons commands sending several carriage returns will cause the Storage Module to match the computer s baud rate see specifications above 1 4 6 File Marks File Marks are used to separate data in the Stor...

Страница 12: ...The delay increases by approximately one second for each increase in the address number The first flash after the delay will last from 1 to 2 seconds depending on the last use of the module Further fl...

Страница 13: ...of the flash memory adequate for most purposes An SM16M will take around 3 minutes to perform the quick reset 1 5 2 Programming the Datalogger to Send Data to the Storage Module Instruction 96 is use...

Страница 14: ...Section 1 SM4M SM16M Storage Module Overview 1 8 This is a blank page...

Страница 15: ...e with address X X 1 8 30 21X and CR7 use option code 30 to send data to the Storage Module When output to the Storage Module is enabled with Instruction 96 the Storage Module s may either be left on...

Страница 16: ...ser should connect the SM to the datalogger and wait for the green LED to light When the LED goes off data transfer is complete and the SM can be disconnected To avoid data loss data must be retrieved...

Страница 17: ...cution of multiple instances of Instruction 96 to Storage Modules Add 5ms to the time it takes to store the number of bytes to be transmitted 1ms per byte before executing the next Instruction 96 whic...

Страница 18: ...n a unique format which includes calibration measurements at the beginning and end of the series of measurements Split part of the PC208 PC208W software packages will translate this Burst data 2 3 1 B...

Страница 19: ...the baud rates of the Storage Module and the datalogger are matched The SM4M SM16M will only accept data at 9600 or 76800 baud There is no user configuration for this setting in the module as it autom...

Страница 20: ...imes more memory space than binary format Consider buying an operating system upgrade for dataloggers that do not allow binary data storage with P96 3 2 3 9 Mode Keyboard Initiated Data Transfer The 9...

Страница 21: ...re data at 76 800 baud When the Burst Measurement Instruction 23 is used with the 21X select the destination parameter 4 option as 2 for serial port output at 76 800 baud to send data to the Storage M...

Страница 22: ...Section 3 The Printer Enable Method of Data Storage 3 4 This is a blank page...

Страница 23: ...r more to be certain it places a File Mark in the Storage Module when reconnected A File Mark can be placed in the data by using SMS the 9 Mode command 3 not in 21X or CR7 or from a special P96 comman...

Страница 24: ...In this case the byte will be stored with a null character ASCII code 0 as the second byte If a single byte remains in the input buffer when power is disconnected from the Storage Module it will be s...

Страница 25: ...ing the data This process does have some consequence for memory use Data is stored as received With flash memory it is not possible to overwrite the previously received data transmission without erasi...

Страница 26: ...ocks The next block containing the remainder of the bad data has not been erased In both cases the bad data will be collected as normal data and written to a file However when the end of bad data mark...

Страница 27: ...respectively to activate the Storage Module Section 5 2 SMS performs this function automatically 9227 5 H a a Logan UT SM4M SN SOLID STATE STORAGE MODULE MADE IN USA FIGURE 5 1 Storage Module Connect...

Страница 28: ...de is limited to a maximum speed of 9600 baud 19200 on the CR23X which can make the transfer of large volumes of data a lengthy process compared to direct connection to a PC where baud rates up to 115...

Страница 29: ...1 for direct connection For communication via the CR500 510 CR10 10X or CR23X the setup shown in Figure 5 2 is required The CR23X can also be used without an SC32A Figure 5 3 SMS also has a terminal...

Страница 30: ...e Module address 1 8 If the connections are correctly made and the Storage Module is properly addressed it will send back CR LF If the attempt to establish communications fails the datalogger will ret...

Страница 31: ...wer up This feature is standard in the CR500 510 CR10 10X and CR23X dataloggers The program should be stored in program number eight in the module with address one 21X and CR7 PROMs before the current...

Страница 32: ...commands or by manually deleting each of the programs in the module In the latter case after the last program is erased the module will proceed to erase and tidy up the program storage area The 21X o...

Страница 33: ...SM A 01 0000 Enter command 1 Ready for instruction to effect command 1 248 01 248 Key in 248 to effect command 1 A 01 Execute SM Reset The module then executes the equivalent of the 1249K telecommuni...

Страница 34: ...inserted 0 indicates that it was not 4 04 XX DISPLAY SET MEMORY CONFIGURATION enter the appropriate code to change configuration 0 ring 1 fill and stop 5 DISPLAY STATUS A to advance to each window 01...

Страница 35: ...ay Location Pointer DLP is used to display Storage Module data The Storage Reference Pointer SRP indicates the next location to be written to in the Storage Module The Dump Pointer is an internal Stor...

Страница 36: ...result as 1B Status is displayed following execution of command This command may take some time to complete nnnnC Comma Delineated ASCII Dump Dump nnnn output arrays from the location of the DLP The...

Страница 37: ...designate a new area of memory in the Storage Module File Marks are automatically placed in data at power up upon connection to datalogger or retrieval interface or when a CR500 510 CR10 10X or CR23X...

Страница 38: ...instead A module with damaged memory requires a full memory test to determine how much usable memory remains The switches are not changed to their default settings abfeL Switch Settings a is the Stora...

Страница 39: ...w 0 VDC the processor is set to receive data asynchronously on Pin 9 TD When the PE line is dropped data remaining in the input buffer are stored and the location pointers are saved The processor then...

Страница 40: ...This is a blank page...

Страница 41: ...rsibly erased during the upgrade process To load the operating system you will need to connect the module to a PC running Windows via an SC532 A interface Please follow the instructions supplied with...

Страница 42: ...This is a blank page...

Страница 43: ...can be modified to connect to the datalogger s power terminals as follows 1 Unplug field cable from datalogger and SC532A 2 Cut off DB9 connector 3 Remove cable sheath exposing positive RED wire and...

Страница 44: ...This is a blank page...

Страница 45: ...This is a blank page...

Страница 46: ...campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientifi...

Отзывы: