SDM-CVO4 4-Channel Current/Voltage Output Module
; Wait 100 ms for it to power-on and the outputs to stabilize
; You may need to increase this delay if the sensors ; themselves
; have a power-on delay
3: Excitation with Delay (P22)
1: 1
Ex
Channel
2: 0000
Delay W/Ex (units = 0.01 sec)
3: 10
Delay After Ex (units = 0.01 sec)
4: 0000
mV
Excitation
; Take the four measurements from the Pressure sensors.
4: Volt (Diff) (P2)
1: 4
Reps
2: 04
250 mV Slow Range
3: 1
DIFF
Channel
4: 5
Loc [ Press_1 ]
5: 1.0
Mult
6: 0.0
Offset
; Now turn off the SDM-CVO4 by using the command with 0 reps.
5: SDM-AO4 (P103)
1: 0
Reps
2: 00
SDM
Address
3: 1
Loc [ Scale_1 ]
; Now the normal output processing and final storage
; instructions would follow, for example:
; Every hour set the output flag
6: If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 60
Interval (same units as above)
3: 10
Set Output Flag High (Flag 0)
; Store time
7: Real Time (P77)
1: 1110
Year,Day,Hour/Minute (midnight = 0000)
; Store the average readings
8: Average (P71)
1: 4
Reps
2: 5
Loc [ Press_1 ]
23