background image

Q-7.1 NET RADIOMETER

8

recommends recalibration every six months,
depending on the climate where the sensors
are used.

Each instrument is calibrated by the
manufacturer in a temperature-controlled
calibration chamber by comparison to a
transfer standard.  The chamber uses a
combination of tungsten-halide source and
black body chambers.  The transfer standard is
calibrated by comparison to a precision
pyranometer using a partial shading technique
developed by Radiation Energy Balance
Systems.  The long wave calibration of the
standard has also been verified using black
body chambers.

Care should, however, be taken when
comparing readings with sensors from different
manufacturers or even older models from
Radiation Energy Balance Systems.  This is
because in recent years academic studies have
reported inadequacies both in the design and
calibration of net radiometers from most
manufacturers.  Radiation Energy Balance
Systems aim to keep at the forefront of such
developments and have published corrections
for the calibration of older sensors.  However,
these corrections cannot fully bring the
instruments to the same level of performance
as the new sensors.  It is, however, possible to
upgrade and recalibrate the sensor heads of
older models.

Содержание Q-7.1

Страница 1: ...INSTRUCTION MANUAL Q 7 1 Net Radiometer Revision 5 96 C o p y r i g h t c 1 9 9 1 1 9 9 6 C a m p b e l l S c i e n t i f i c I n c ...

Страница 2: ...er warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorization The following contact information is for US and International customers residing in countries served by Campbell Scientific Inc directly ...

Страница 3: ...te to a vertical or horizontal pipe or rod that is less than 38 mm 1 5 inches in diameter with the two larger U bolts Attach the radiometer support arm to the mounting plate with the two smaller U bolts The radiometer support arm does not need to be level Fasten the cable to the pipe or rod with tape or plastic ties to prevent strain on the wires and damage to the instrument CAUTION Ensure that th...

Страница 4: ...e 500 mV range for the 21X or CR7 The slow integration with 60 Hz rejection yields a more noise free reading For still air net radiation Q can be computed from the thermopile voltage Vt by If Vt 0 Q Wm 2 Vt mV x Fp If Vt 0 Q Wm 2 Vt mV x Fn where Fp and Fn are the positive and negative calibration factors respectively The calibration factors and serial number are given on a label under the sensor ...

Страница 5: ...s using fixed correction factors of 1 045 for positive fluxes and 1 000 for negative fluxes If the wind speed is measured the functions above can be programmed into the datalogger to give a real time correction of net radiation for the effects of wind speed Note that the wind speed is expressed in units of meters per second If the wind speed is measured in units other than meters per second conver...

Страница 6: ...nd input location assignments On the 21X and CR7 use 500 mV input range For a Single Ended measurement use Instruction 1 3 3 PROGRAMMING EXAMPLE WITH DYNAMIC WIND SPEED CORRECTION The following CR10 X program measures an R M Young wind sentry anemometer model 03101 and a Q 7 1 The wind speed is used to give a real time wind correction to net radiation The wind speed must be measured before the win...

Страница 7: ...rections 10 Beginning of Subroutine P85 1 2 Subroutine 2 11 Z X F P37 1 1 X Loc ws_m_s 2 00174 F 3 4 Z Loc A 12 Z X F P34 1 4 X Loc A 2 99755 F 3 7 Z Loc Corr_Fact Enter the negative multiplier n nnn 13 Z X F P37 1 2 X Loc Rn_mV 2 n nnn F 3 3 Z Loc Rn_W_m2 14 Z X Y P36 1 3 X Loc Rn_W_m2 2 7 Y Loc Corr_Fact 3 3 Z Loc Rn_W_m2 15 End P95 Proper entries will vary with program and datalogger channel an...

Страница 8: ...ng Desiccant Make sure that the desiccant is dry i e still blue and white If it has turned pink replace it with dry desiccant or bake it until it is dry see section 4 1 above If the desiccant is pink only at the tube end nearest the sensor head this indicates a leak somewhere on the sensor Windshields and O rings Check that the windshields and O rings are in good condition and are properly seated ...

Страница 9: ... lower windshields are completely seated on the frame around the entire periphery CAUTION Do not attempt to dismantle the instrument body If internal attention is needed contact Campbell Scientific Inc for repair If the windshields become dented remove the dents as follows 1 Remove the vinal cap from the support arm 2 Blow gently into the end plug breather hole to inflate the windshields Remember ...

Страница 10: ...f the standard has also been verified using black body chambers Care should however be taken when comparing readings with sensors from different manufacturers or even older models from Radiation Energy Balance Systems This is because in recent years academic studies have reported inadequacies both in the design and calibration of net radiometers from most manufacturers Radiation Energy Balance Sys...

Страница 11: ...This is a blank page ...

Страница 12: ... campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Ltd CSL Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UNITED KINGDOM www campbellsci co uk sales campbellsci co uk Campbell Scientific Ltd France Miniparc du Verger Bat H 1 rue de Te...

Отзывы: