Campbell OXYGUARD TYPE III CS512 Скачать руководство пользователя страница 5

Campbell Scientific CS512 Oxyguard 
Type III Dissolved Oxygen Probe 

An accurate and reliable sensor is a critical element in any measurement system.  The 
Oxyguard Stationary Probe meets these criteria for the measurement of dissolved oxygen. 

1.  General Information 

The Oxyguard Stationary Probe is a galvanic probe which produces a millivolt 
signal proportional to the amount of oxygen present in the measured medium.  
Oxygen diffuses through the membrane onto the cathode, reacts chemically, and 
combines with the anode.  An electrical current is produced by this chemical 
reaction which is converted from microamps to millivolts by an in-line resistor.  
An in-line thermistor also conditions the signal providing automatic temperature 
compensation.  With these features, the probe produces a millivolt output 
proportional to the oxygen present in the medium in which it is placed.  The probe 
consists of two parts, an upper part with cathode, anode and cable, and a lower part 
comprising a screw on membrane cap with fitted membrane.  The cap is filled with 
electrolyte and simply screwed onto the top component. 

The probe is self-polarizing and requires no external power source.  There are two 
wires to connect.   

Because the probe's output is linear, it is possible to connect it directly to a data 
acquisition system capable of handling the small millivolt signal. 

The probe's robust construction and simple design make maintenance and 
servicing it straightforward.  There is no need to send the probe back to the factory 
for servicing.  It utilizes a strong, easy-to-clean and easy-to-change membrane in a 
screw-on membrane cap.  Regular servicing is not required.  When necessary the 
probe can be fully overhauled in five minutes. 

 

FIGURE 1-1.  CS512 

Содержание OXYGUARD TYPE III CS512

Страница 1: ...Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe Revision 12 06 C o p y r i g h t 2 0 0 1 2 0 0 6 C a m p b e l l S c i e n t i f i c I n c...

Страница 2: ...u of all other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or cons...

Страница 3: ...formation 1 2 Specifications 2 3 Optional Probe Accessories 3 4 Optional Agitator 3 5 Application Information 4 6 Wiring 4 7 Programming 5 8 Calibration 6 9 Maintenance 7 10 Agitator Control 7 Appendi...

Страница 4: ...P2 5 3 Sample Program using VoltSE 5 4 Sample Program using VoltDiff 5 5 Sample Program using VoltSE 5 6 Sample Program using P36 6 7 Sample Program using Expression 6 8 Sample Program using P86 and...

Страница 5: ...the probe produces a millivolt output proportional to the oxygen present in the medium in which it is placed The probe consists of two parts an upper part with cathode anode and cable and a lower par...

Страница 6: ...rin Membrane O ring Buna N Body Seal O ring Viton 0 05 mm 2 mil Teflon Dimensions and Weight 2 28 diameter x 2 3 height 15 9 oz Cable standard length 7 m 23 ft other lengths available upon request 2 w...

Страница 7: ...es the water also has a high bio loading and the probes become fouled resulting in inaccurate DO measurements FIGURE 4 1 Preventing Bio fouling of the CS512 The PT4 Agitator overcomes these problems T...

Страница 8: ...plication Information SCWin users This manual was written primarily for those whose needs are not met by SCWin Your procedure is much simpler just add the CS512 probe it s in the water folder save you...

Страница 9: ...60 Hz Rejection Range code 23 used for CR23X 3 1 SE Channel 4 1 Loc DOmV 5 1 0 Multiplier See Calibration 6 0 0 Offset EXAMPLE 2 Sample Program using P2 1 Volt Diff P2 1 1 Reps 2 24 250 mV 60 Hz Reje...

Страница 10: ...ing a multiplier of one 6 Change the multiplier in the datalogger program from one to the calculated number A more common way to enter the multiplier is to insert a separate instruction in the program...

Страница 11: ...peat cycle timer requires no datalogger programming However some users choose to use a solid state relay and have the datalogger agitate the water on the probe face either periodically throughout the...

Страница 12: ...et instruction EXAMPLE 9 Sample Program using Portset Portset 1 1 Delay 500 msec Portset 1 0 In the CR1000 CR800 CR850 and CR3000 dataloggers use the Portset instruction EXAMPLE 10 Sample Program usin...

Страница 13: ...A 1 Appendix A Dissolved Oxygen in Fresh Water Table...

Страница 14: ...Appendix A Dissolved Oxygen in Fresh Water Table A 2...

Страница 15: ...This is a blank page...

Страница 16: ...campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientif...

Отзывы: