CS510 OxyGuard Dissolved Oxygen Probe

6

3)  Determine the air temperature and barometric pressure.

4)  Using a calibration chart such as that provided in the probe’s manual,

determine the oxygen concentration of the air.

5)  Use the following equation to calculate the multiplier:

M = P/R
M = Multiplier
P = Concentration in PPM of the air (from the calibration  chart)
R = The signal output of the OxyGuard probe when using a multiplier of one

6)  Change the multiplier in the CR10 program from one to the calculated

number.

10.  Maintenance

The OxyGuard probe needs little maintenance. Regular cleaning of the
membrane is all that is required. The membrane is very durable and can be
cleaned with a cloth or soft paper. Do not scratch it clean with your fingernail.
Please see the OxyGuard manual for complete maintenance procedures.

11.  Agitator Control

In low flow conditions (less than about 2 cm/sec), it may be necessary to add
an agitator to the OxyGuard probe. The distributor for OxyGuard in North
America sells a self-timed agitator which can be purchased through Campbell
Scientific. The self-timed agitator requires no datalogger programming.
However, some users choose to use a solid state relay and have the datalogger
agitate the water on the probe face either periodically throughout the day or
just before measurement. Agitating just before the measurement saves on
power and causes less wear and tear on the agitator and probe membrane.

The wiring for the agitator as controlled by this example program would be as
follows:

WHITE

BLACK

Agitator

 +                                   -

12 V Battery

4                                      1

#7321 Relay

Crydom D1D07

3                                      2

Ground/G

Datalogger

Port C1

Содержание Oxyguard CS510

Страница 1: ...CS510 OXYGUARD DISSOLVED OXYGEN PROBE INSTRUCTION MANUAL 1 00 COPYRIGHT c 2000 CAMPBELL SCIENTIFIC INC ...

Страница 2: ...This is a blank page ...

Страница 3: ... of sale and return policy apply to these items Products may not be returned without prior authorization To obtain an RMA number contact Point Four Systems Inc phone 604 936 9936 NOTE if the product has been exposed to hazardous media it must be fully decontaminated and neutralized prior to return to Point Four Systems ...

Страница 4: ...This is a blank page ...

Страница 5: ...xygen present in the medium in which it is placed The probe consists of two parts an upper part with cathode anode and cable and a lower part comprising a screw on membrane cap with fitted membrane The cap is filled with electrolyte and simply screwed onto the top component The probe is self polarizing and requires no external power source There are only two wires to connect Cable length is rarely...

Страница 6: ...gs Viton Membrane 0 05 mm 2 mil Teflon Dimensions and Weight 58 mm diameter by 59 mm high 2 1 4 inch diameter by 2 1 4 inch high 350 g 3 4 lb including standard 3 m 10 ft cable length Cable 2 conductor 0 75 square mm 16 gauge standard length 3 m 10 ft other lengths available upon request Operating Conditions Temperature 0 to 50 C 32 to 122 F Pressure Maximum 10 atmospheres 147 psig Minimum Submers...

Страница 7: ... accurate calibration 5 Optional Agitator The PT4 Agitator is a reliable and robust agitator for use in conjunction with probes subjected to bio fouling in ponds and stagnant water conditions O2 probes require a minimum water velocity across their membranes to function properly Therefore to measure DO in stagnant water conditions it is necessary to move the water past the membrane to get accurate ...

Страница 8: ...elocity as well as prevents the membrane from becoming bio fouled The optimum sweeping frequency depends upon the design of probe and type of membrane used and water conditions An ON time of 0 25 seconds and OFF time of 5 seconds is suitable in most circumstances Agitator s overall size 7 1 8 x 3 1 4 diameter 1 1 4 lb 180mm x 83mm 0 6 kg Supplied with 10 ft 3 meters cable Power required 10 5 to 18...

Страница 9: ...t 7 Wiring The OxyGuard probe uses one differential channel Differential wiring is better at rejecting electrical noise and ground loop error Red 1H Blue 1L 8 Programming Instruction 2 differential millivolt measurement is required for the OxyGuard probe The example measurement instructions that follow do not store data to final storage Additional instructions Typically P92 P77 and output processi...

Страница 10: ...not scratch it clean with your fingernail Please see the OxyGuard manual for complete maintenance procedures 11 Agitator Control In low flow conditions less than about 2 cm sec it may be necessary to add an agitator to the OxyGuard probe The distributor for OxyGuard in North America sells a self timed agitator which can be purchased through Campbell Scientific The self timed agitator requires no d...

Страница 11: ...er as wired 2 Excitation with Delay P22 1 1 Ex Channel 2 0 Delay W Ex units 0 01 sec 3 25 Delay After Ex units 0 01 sec 4 0 mV Excitation 3 Do P86 1 51 Set Port 1 Low The above example is not as power efficient as possible and would require AC power If it is necessary to operate an agitator without AC power available write the program so that the agitator is only operated for a short period of tim...

Страница 12: ...CS510 OxyGuard Dissolved Oxygen Probe 8 This is a blank page ...

Отзывы: