
OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
28
SSC = Wt
s
[V
w
+ Wt
s
/ρ
s
]
–1
, where Wt
s
= total sediment weight in tub in mg,
V
w
= volume of water in liters, ρ = density of water (ρ = 1.0 kg L
–1
at 10 °
C),
and ρ
s
= sediment density (assume 2.65 10
3
mg L
–1
)
Procedure
1.
Record and log the clean-water signal as in Step 1, Section
(p. 22)
. Use the same value, such as, sidescatter,
backscatter, or ratio throughout the calibration.
2.
Move the OBS500 to the suspender as described in setup.
3.
Weigh 500 ± 10 mg of sediment in a weigh boat and transfer it to the
teacup. Record the weight on the calibration log sheet and add about 10 cc
of water from the suspender tub to the teacup and mix the water and
sediment into a smooth slurry with the teaspoon.
4.
Add the sediment slurry to the tub and rinse the teacup and spoon with tub
water to get all the material into the suspender.
5.
Turn the suspender on and let it run for 10 minutes or until the OBS signal
stabilizes.
6.
Take averages of signals with the computer or datalogger and enter them
on the calibration log sheet.
7.
Calculate the sediment-weight increment as follows: W
i
= 2500 mg
(4000/V
x
), where W
i
= the incremental weigh of sediment and V
x
= the
average output signal from step 6. The resulting weight gives the amount
of sediment to add in order to have evenly spaced calibration points.
8.
Add enough additional sediment to get one full increment of sediment, W
i
± 5%. Repeat steps 4, 5, and 6.
9.
Repeat step 8 until five full increments of sediment have been added or
until the OBS signals exceed the output range.
10.
Perform 3
rd
order polynomial regressions on the data to get the coefficients
for converting OBS output to SSC.
7.6 Programming
Short Cut
is the best source for up-to-date datalogger programming code.
Programming code is needed when:
•
Creating a program for a new datalogger installation
•
Adding sensors to an existing datalogger program
If your data acquisition requirements are simple, you can probably create and
maintain a datalogger program exclusively with
Short Cut
. If your data
acquisition needs are more complex, the files that
Short Cut
creates are a great
source for programming code to start a new program or add to an existing
custom program.
Содержание OBS500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv ...
Страница 15: ...OBS500 Smart Turbidity Meter with ClearSensor Technology 5 FIGURE 4 4 Place and secure mounting fixture ...
Страница 58: ...Appendix B Example Programs B 6 ...
Страница 60: ...Appendix C OBS500 Copper Sleeve Kit Installation C 2 ...
Страница 63: ......