OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
25
FIGURE 7-6. Position of OBS500 in clean tap water in big black tub
2.
Pour the first SDVB standard into the appropriately sized cup (see TABLE
).
3.
Position the OBS sensor in the cup as shown in FIGURE
10-second averages of the low- and high-range outputs. Record the
average outputs on the calibration log sheet.
4.
Pour the standard back into its container.
5.
Wipe sensor with a clean, dry towel to remove residual standard.
6.
Repeat steps 2, 3, 4, and 5 for the other standards.
7.
Perform 2
nd
-order polynomial regressions on the calibration data to get the
coefficients for converting OBS signals to TU values.