OBS-5+ System
adjustment takes two seconds. If your PC is set to the wrong rate for some
reason, use the check box to select
ONLY change host computer port
. Then
click
Apply
and the
OBS Settings
button.
If you get the OBS-5+ information box, the baud rate of the unit is
synchronized with your PC. If you don’t get an information box, repeat the
above procedure until communication is established.
6.2.5 Testing Sensors
1)
Before daily operations and deployments, verify that the instrument works
by clicking
Open Plot,
and then clicking
Survey
. Select all
sensors and click
Start Survey
. An example plot of data is shown below.
12
Содержание OBS-5+