OBS-3+ and OBS300 Suspended Solids and Turbidity Monitors 

TABLE 9-1.  Troubleshooting Chart 

Fault 

Cause of Fault 

Remedy 

Fails finger wave test 

No power, dead battery 

Replace battery and reconnect wires 

 

MCIL-5 plug not fully seated 

Disconnect and reinsert plug. 

 

Sensor broken 

Visually inspect for cracks.  Return 

the sensor to manufacturer if cracks 

are found. 

 

Electronic failure.  Units draws less than 11 

mA or more than 40 mA. 

Return the sensor to manufacturer. 

Fails shake test 

Sensor leaked 

Return the sensor to manufacturer. 

Fails calibration check 

Aging of light source causes it to get dimmer 

with time. 

Recalibrate (see Section 8, 

Calibration

 

10.  Maintenance 

There are no user-serviceable parts inside the sensor 

housing.  Do not remove the sensor or connector from 

the pressure housing.  This will void the warranty and 

could cause a leak. 

The most important maintenance item is keeping the window clean.  A  

Scotch-Brite

 scouring pad works well for most types of window fouling.  

First wet the pad and then place it on a counter with a plastic-laminate top so 

that the side of the pad is aligned with the edge of the counter.  Work the 

window of the OBS sensor back and forth on the pad until it is clean while 

removing as little epoxy as possible.  If encrusting organisms such as barnacles 

or tube worms have attached to the sensor, it will have to be gently scraped 

with a flexible knife blade prior to using the pad.  Some applications will result 

in pitting of the sensor face.  Pits can be removed with abrasive cloth.  Polish 

the sensor window as follows:  

1.

 

Tape a strip of 400 grit wet-or-dry abrasive cloth to the edge of a 

counter (see above). 

2.

 

Add a few drops of water to the abrasive and work the sensor window 

in smooth one-way strokes on the cloth using the counter edge as a 

guide. 

3.

 

Continue until the sensor is shiny and pit free.   

It is important to remove as little epoxy as possible. 

 

Do not use solvents such as MEK, toluene, acetone, or 

trichloroethylene on OBS sensors. 

WARNING 

NOTE 

WARNING 

22 

Содержание OBS-3+

Страница 1: ...INSTRUCTION MANUAL OBS 3 and OBS300 Suspended Solids and Turbidity Monitors Revision 9 14 C o p y r i g h t 2 0 0 8 2 0 1 4 C a m p b e l l S c i e n t i f i c I n c...

Страница 2: ......

Страница 3: ...ucts to CSI CSI will return such Products by surface carrier prepaid within the continental United States of America To all other locations CSI will return such Products best way CIP port of entry per...

Страница 4: ...g container Campbell Scientific s shipping address is CAMPBELL SCIENTIFIC INC RMA _____ 815 West 1800 North Logan Utah 84321 1784 For all returns the customer must fill out a Statement of Product Clea...

Страница 5: ...ore beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit...

Страница 6: ......

Страница 7: ...urement Details 7 6 Specifications 8 7 Installation 9 7 1 Pre Deployment Tests 9 7 2 Mounting Considerations 10 7 2 1 Siting 10 7 2 2 Mounting Options 10 7 2 2 1 PVC Pipe 10 7 2 2 2 Cable Ties or Hose...

Страница 8: ...B 1 CRBasic Examples B 1 B 1 1 CR1000 Example Program B 1 B 1 2 CR200 X Example Program B 2 B 2 Edlog Example B 3 C Electrical Connections Details C 1 D Datalogger Connection to a Relay D 1 Figures 5...

Страница 9: ...sor to a wide range of SSC 26 11 5 Infrared reflectivity of minerals as a function of 10 Munzell value 27 C 1 Pin assignments for MCBH and MCIL wet pluggable connectors C 1 D 1 Wiring diagram for conn...

Страница 10: ...Table of Contents iv...

Страница 11: ...ific instruments The titanium body option option TB must be used if submersing the probe in seawater Using an OBS sensor with a stainless steel housing option SB in seawater voids the warranty and cau...

Страница 12: ...nd mounting options 4 2 Use SCWin to Program Datalogger and Generate Wiring Diagram Short Cut is an easy way to program your datalogger to measure the sensor and assign datalogger wiring terminals The...

Страница 13: ...S 3 and OBS300 Suspended Solids and Turbidity Monitors 3 When Short Cut opens select New Program 4 Select Datalogger Model and Scan Interval default of 5 seconds is OK for most applications Click Next...

Страница 14: ...tage in units of Enter the Low Range Coefficients and High Range Coefficients according to your calibration sheet see Section 7 5 Calibration Certificate 6 After selecting the sensor click at the left...

Страница 15: ...organic matter Ambient light rejection and low temperature coefficient 5 1 Applications OBS sensors are used for a wide variety of monitoring tasks in riverine oceanic laboratory and industrial settin...

Страница 16: ...t suspended matter in water and turbidity from the relative intensity of light backscattered at angles ranging from 90o to 165o in clean water A 3D schematic of the main components of the OBS 3 is sho...

Страница 17: ...6 Specifications It is also possible to purchase sensors configured to operate from 5 V power however the output span is limited to 2 5 V The sensor needs to be connected to a datalogger current mete...

Страница 18: ...0 X CR23X 21X and CR7 Operating Temperature 0 to 40 C Sensor may be damaged if it is encased in frozen liquid Ranges Turbidity low high 250 1000 NTU 500 2000 NTU 1000 4000 NTU Mud1 5000 to 10 000 mg L...

Страница 19: ...field installations 1 Connect the red and black wires in the cable supplied with the sensor to a 9 or 12 Vdc battery 2 Plug the cable into the sensor and connect a multimeter across the blue test lea...

Страница 20: ...e OBS 3 OBS300 will fit inside a 1 25 in schedule 40 PVC pipe 7 2 2 2 Cable Ties or Hose Clamps The OBS 3 OBS300 can be mounted to a frame or post using large high strength nylon cable ties 7 6 mm 0 3...

Страница 21: ...10X White High Range Signal Single ended Input Single ended Input Blue Low Range Signal Single ended Input Single ended Input Green Signal Ground AG Black Power Ground G G Red Power SW12V SW12V Clear...

Страница 22: ...low input range The millivolt measurements are converted to NTU by using the coefficients provided on the Calibration Certificate We supply a calibration curve in units of both millivolts and volts se...

Страница 23: ...3 and OBS300 Suspended Solids and Turbidity Monitors FIGURE 7 2 Calibration certificate showing millivolt coefficients Curve Type Polynomial for Converting to NTU Polynomial for Converting to NTU CSL...

Страница 24: ...ene divinylbenzene beads SDVB turbidity over the range from 0 to 4000 NTU is shown on FIGURE 8 1 As shown on the inset the response function is contained within region A the linear region of the unive...

Страница 25: ...meter cannot be used to calibrate a different model The standard values for calibrating OBSs are 125 NTU GFS Part No 8428 250 NTU GFS Part No 8429 500 NTU GFS Part No 8430 1000 NTU GFS Part No 8431 20...

Страница 26: ...the dilution procedures the uncertainty in the NTU value of an SDVB standard is 1 of the value indicated on the standard bottle Consequently the NTU value of one liter of standard in an uncovered 100...

Страница 27: ...calibration log sheet 2 Swap the DMM or datalogger test lead to the high range output lead white and record a 10 second average of the high range output Record the average output on the calibration lo...

Страница 28: ...n which it is suspended can be accurately measured with volumetric glassware Of the three methods dry sediment calibration causes the greatest physical and chemical alteration of the sediment Alterati...

Страница 29: ...the OBS sensor will be used sediment should be in a state where grinding sieving or pulverization does not change its particle size distribution Averaging DMM or datalogger with test leads 12 V gel c...

Страница 30: ...ransfer it to the teacup Record the weight on the calibration log sheet and add about 10 cc of water from the suspender tub to the teacup and mix the water and sediment into a smooth slurry with the t...

Страница 31: ...r there is a problem that requires attention 2 The Shake Test is done to determine if water has leaked inside the pressure housing Unplug the cable and gently shake the sensor next to your ear and lis...

Страница 32: ...an A Scotch Brite scouring pad works well for most types of window fouling First wet the pad and then place it on a counter with a plastic laminate top so that the side of the pad is aligned with the...

Страница 33: ...ize shape and color remaining constant particle area normal to a light beam will determine the intensity of light scattered by a volume of suspended matter Results of tests with sediment shown on FIGU...

Страница 34: ...is to take a lot of water samples and analyze them in the laboratory This is not always practical during storms when the errors are likely to be largest Do not rely solely on OBS sensors to monitor su...

Страница 35: ...e particle and the detector that backscatter decreases exponentially with increasing sediment concentration For mud this occurs at concentrations greater than about 5000 mg L 1 FIGURE 11 4 shows a cal...

Страница 36: ...tests have shown that whiteness color and IR reflectivity are correlated Calcite which is highly reflective and white in color will produce a much stronger OBS signal on a mass concentration basis tha...

Страница 37: ...7 Bubbles and Plankton Although bubbles efficiently scatter light monitoring in most natural environments shows that OBS signals are not strongly affected by bubbles Bubbles and quartz particles backs...

Страница 38: ...ic sells two wipers from a third party manufacturer the Hydro Wiper C with its own controller or the Hydro Wiper D that is controlled by a datalogger 12 References Bohren C F and D H Huffman Absorptio...

Страница 39: ...sample the environment or measurement procedure that produce unintended and unknown shifts in turbidity or SSC Sample Volume The water volume where the OBS infrared beam and suspended particles inter...

Страница 40: ...o will water turbidity and as the concentration of dissolved light absorbing matter increases turbidity will decrease Turbidity operational definition NTU value is a number ranging from 0 to 10 000 th...

Страница 41: ...for editing A 1 1 CRBasic Datalogger Use the following procedure to import Short Cut code into CRBasic Editor CR200 X CR1000 CR800 CR3000 CR5000 CR9000 X dataloggers 1 Create the Short Cut program fol...

Страница 42: ...3 Click File Document DLD File Assuming the default paths were used when Short Cut was installed navigate to C CampbellSci SCWin folder The file of interest has a DLD extension Select the file and cl...

Страница 43: ...Public A 2 B 2 C 2 i n NTUX 21 Define Data Tables DataTable Data_15m 1 1 DataInterval 0 15 Min 10 Sample 1 NTU FP2 EndTable Main Program BeginProg Declare Constants n 21 A 1 0 24763 B 1 105 69 C 1 0 2...

Страница 44: ...Table Data_15m NextScan EndProg B 1 2 CR200 X Example Program The following example uses the voltage curve of the calibrations certificate provided in FIGURE 7 3 Since the coefficients of the voltage...

Страница 45: ...t 0 CallTable NTUData NextScan EndProg B 2 Edlog Example The following example uses the voltage curve of the calibrations certificate provided in FIGURE 7 3 Since the coefficients of the voltage curve...

Страница 46: ...Loc NTU_V_1 3 2 Avg Loc NTU_Volts NTU 0 29928 105 69 NTU_Volts 0 24763 NTU_Volts 2 evaluate to see if value 250 where High range measurement should be taken via Sub1 7 If X F P89 1 1 X Loc NTU 2 3 3...

Страница 47: ...Appendix B Example Programs NTU 0 20709 401 11 NTU_Volts 35 310 NTU_Volts 2 6 End P95 End Program B 5...

Страница 48: ...Appendix B Example Programs B 6...

Страница 49: ...concerned with the wire colors for the 8425 cable as the MCBH wires are not accessible When a custom cable assembly is purchased from a third party vendor to connect the OBS sensor to a current meter...

Страница 50: ...BS sensor bulkhead connectors MCBH 5 FS MCIL 5 MP Contact Number Electrical Function Wire Color MCBH Wire Color 8425 1 Power 5 15V Red Red 2 Power Ground Black Black 3 Signal Common Green Green 4 High...

Страница 51: ...e Green FIGURE D 1 Wiring diagram for connecting an OBS sensor to an external relay and a datalogger 1 The assignment of channel number for example SE Channel 1 or C1 may vary depending on the applica...

Страница 52: ...Appendix D Datalogger Connection to a Relay D 2...

Страница 53: ......

Страница 54: ...m br Campbell Scientific Canada Corp CSC 14532 131 Avenue NW Edmonton AB T5L 4X4 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Centro Caribe S A CSCC 300 N Cementerio Edific...

Отзывы: