LI200X Pyranometer
2. Place the LI200X in the center of the LI2003S base/leveling fixture.
3. Loosely mount the LI2003S base/leveling fixture on the 015ARM. Do not
fully tighten the three mounting screws.
4. Turn the leveling screws as required to bring the bubble of the bubble level
within the ring.
5. Tighten the mounting screws to secure the assembly in its final position.
Check that the pyranometer is still correctly leveled and adjust as
necessary.
6. Route the sensor cable along the underside of the 015ARM’s arm to the
tripod/tower, and to the instrument enclosure.
7. Secure the cable to the mounting arm and mast using cable ties.
8. Remove the red cap after installing the sensor. Save this cap for shipping
or storing the sensor.
Bubble Level
Holes for
Mounting Screws
LI2003S Base
Leveling
Screws
Bubble Level
LI200X
Leveling Screws
Cable Tie
LI2003S
015ARM
Mounting Screws
9