Appendix C. EasyFlux
®
DL CR6KH20
C-2
The VOLT116 and CDM-A116 are functionally the same,
however their OSes are not interchangeable. If updating an OS,
make sure it is for the correct model.
C.2 Precautions
EasyFlux DL CR6KH20
requires the CR6 to have operating system (OS)
version 09.02 or newer. If using a VOLT116, it must have OS v.01 or
newer, or if using a CDM-A116, it must have v.06 or newer.
The program applies the most common EC corrections to fluxes. However, the
user should determine the appropriateness of the corrections for their site.
Campbell Scientific always recommends saving time-series data in the event
reprocessing of raw data is warranted. Further, the user should determine
the quality and fitness of all data for publication, regardless of whether
said data were processed by
EasyFlux DL CR6KH20
or another tool.
As
EasyFlux DL CR6KH20
is not encrypted, users have the ability to view and
edit the code. However, Campbell Scientific does not guarantee the
function of an altered program.
C.3 Wiring
When wiring the sensors to the data logger or VOLT116, the default wiring
schemes, along with the number of instruments
EasyFlux DL CRKH20
supports, should be followed if the standard version of the program is being
used. TABLE
through TABLE
A KH20 and CSAT3B are the only required sensors for the program. The
additional sensors described in the following tables are optional, although the
CS106 and FW3 are recommended. Many of the optional sensors are wired to a
VOLT116 (or CDM-A116) module, which effectively increases the CR6
analog terminals. If one or more of the optional sensors are not used, the data
logger or VOLT116 terminals assigned to those sensor wires should be left
unwired.
If the standard data logger program is modified, the wiring
presented in TABLE
may no longer apply. In these cases,
refer directly to the program code to determine proper wiring.
If using an analog expansion module, all wiring and connections
are the same whether using a VOLT116 or a CDM-A116.
Therefore, throughout this appendix, the wiring terminals are only
listed for the VOLT116.
NOTE
NOTE
NOTE