1.2 HydroSense Modes
The
Water Content Measurement Mode
displays the measurement result as
percent volumetric water content and shows the period of the probe output in
milliseconds.
In the
Water Deficit Mode,
HydroSense measurements are taken at lower and
upper water contents chosen by the user and stored in memory as reference
values. The reference values are then applied to subsequent measurements to
determine the amount of water that must be added to bring the soil to the upper
water content. The relative water content based on the reference values is also
displayed.
See sections 4 and 5 for detailed operating information.
2. Specifications
HydroSense Display Unit
Measurement
% Volumetric Water Content
Resolution
1.0%
Environmental
Splash resistant
Display
16 character, 2 line LCD
Keypad
Two button membrane keypad
Power
2 x AAA 1.5V DC alkaline batteries
Current Consumption
Sleep
<50
µ
A
Idle
1.5mA
Measurement
70mA
Battery Life
Approximately 12 months
Dimensions
120 (w) x 73 (d) x 24 (h) mm
Weight
166 gm
CS620 Water Content Reflectometer
Accuracy
+/- 3.0% volumetric water content with electrical
conductivity <2 dS m-1.
Range
0% to saturation*
Stabilization time
Instantaneous
Response time
< 50 milliseconds
Power requirement
100 milliamps at 5
±
0.03 volts
Output
Square wave with +/- 2.5V DC amplitude
Dimensions
Head
105 x 70 x 18 mm
Rods
120 or 200 mm length , 5 mm diameter, 32
mm spacing