Model HFP01 Soil Heat Flux Plate
3
6.
In
Output Setup
, type the scan rate, meaningful table names, and
Data
Output Storage Interval
.
7.
Select the output options.
8.
Click
Finish
and save the program. Send the program to the data logger if
the data logger is connected to the computer.
9.
If the sensor is connected to the data logger, check the output of the sensor
in data display in
LoggerNet
,
RTDAQ
,
PC400
, or
PC200W
to make sure it
is making reasonable measurements.
5. Overview
The HFP01 Soil Heat Flux Plate uses a thermopile to measure temperature
gradients across its plate. Operating in a completely passive way, it generates a
small output voltage that is proportional to this differential temperature.
Assuming that the heat flux is steady, that the thermal conductivity of the body
is constant, and that the sensor has negligible influence on the thermal flow
pattern, the signal of the HFP01 is directly proportional to the local heat flux.