17
To configure the SC105 using Device Configuration Utility, first connect the
female-female null modem cable supplied with the SC105 to your PC’s serial
port and the Modem connector on the SC105. Next, provide power to the
SC105 by connecting the SC12 cable supplied to the SC105’s Datalogger
connector and the CS I/O port of a datalogger. Once this is done, open Device
Configuration Utility, expand the Peripheral section if available and select
SC105. Check that the Communications Port selected is the one that you have
connected to the SC105, press the green program button on the SC105, then
press Connect in Device Configuration Utility.
FIGURE 9-3. SC105 deployment configuration
To set up the SC105 for use in the modem kit, configure as below.
TABLE 9-3. SC105 Configuration Settings
Setting
Description
CS I/O Mode
Modem Enable
CS I/O Baud Rate
115.2k
RS-232 Mode
PC/PDA
Baud Rate
115.2k