CS230 Temperature Profiler
10
TABLE 7-4. SDI-12 Addresses and Positions
Numeric Set Uppercase Set
Lowercase Set
1 / 1
A / 10
a / 36
2 / 2
B / 11
b / 37
3 / 3
C / 12
c / 38
4 / 4
D / 13
d / 39
5 / 5
E / 14
e / 40
6 / 6
F / 15
f / 41
7 / 7
G / 16
g / 42
8 / 8
H / 17
h / 43
9 / 9
I / 18
i / 44
J / 19
j / 45
K / 20
k / 46
L / 21
l / 47
M / 22
m / 48
N / 23
n / 49
O / 24
o / 50
P / 25
p / 51
Q / 26
q / 52
R / 27
r / 53
S / 28
s / 54
T / 29
t / 55
U / 30
u / 56
V / 31
v / 57
W / 32
w / 58
X / 33
x / 59
Y / 34
y / 60
Z / 35
a / 61
{ – reset to factory address
7.2.2 Slow Sequence Program Instructions
Use of the slow sequence program instructions should be considered if the
CS230 measurement will exceed the program scan interval of the additional
instruments included in the station. For example, if a CS230 consists of 17 or
more temperature sensors, the time required to poll all sensors and receive data
back can be greater than 5 seconds based on the 300 ms execution time for the
aR0!
command. For more details on the use of the slow sequence program
instructions, please reference the related
LoggerNet Help
or relevant datalogger
manual.