This watchdog.txt file need not cause concern; it simply indicates that the program
automatically resumed.
B.1 Using
Device Configuration Utility
1. In
Device Configuration Utility
, connect to the CRVW3 and choose the Terminal tab.
2. Click in the screen and press Enter several times. You should get the CRVW> terminal
prompt.
3. Type help and press Enter. A list of possible terminal commands will be given.
4. Type a command name, along with any required arguments, and press Enter.
NOTE:
If you wait too long between entering commands, you will need to press Enter several
times again to restore the terminal session.
5. When finished with the terminal session, click Close Terminal.
B.2 Using
LoggerNet
1. In the
LoggerNet
Connect Screen, connect to the CRVW3.
2. Select Datalogger menu > Terminal Emulator….
3. Click Open Terminal.
4. Click in the screen and press Enter several times. You should get the CRVW> terminal
prompt.
5. Type help and press Enter. A list of possible terminal commands will be given.
6. Type a command name, along with any required arguments, and press Enter.
NOTE:
If you wait too long between entering commands, you will need to press Enter several
times again to restore the terminal session.
7. When finished with the terminal session, click Close Terminal.
CRVW3 3-Channel Vibrating-Wire Data Logger
76