EESiFlo 6000 Data Logger quick guide 

 

page 1 of 2 

 
 
 
 
 
 

EESiFlo 6000 Series 

Data Logger Setup Guide 

To store data in the internal data logger, activate the EESiFlo 6000 storage function described below. The following 

data can be stored: 

 

Date 

 

Time 

 

Identification of the measuring point 

 

Pipe parameters 

 

Medium parameters 

 

Transducer parameters 

 

Sound path (reflection or diagonal) 

 

Transducer distance 

 

Damping factor 

 

Storage rate 

 

Quantity of measurement 

 

Unit of measurement 

 

Measured values 

 

Values of the totalizers (if setup) 

 

Values of the activated process inputs (if any) 

 

Activating and Deactivating the Storage Function 

 

1.

 

Press the CLR button for 3 seconds and the meter will turn on. 

2.

 

Press the 4 or 6 key to select the OUTPUT OPTIONS program branch then press the Enter button. 

3.

 

Select the channel you wish to setup by pressing the 2 or 8 key then press the Enter button. 

4.

 

Select the Physical Quantity you wish to be displayed by using the 2 or 8 key (Volume flow is the most 
common i.e. GPM) then press the Enter button. 

5.

 

Select the Volume value you wish to be displayed by using the 2 or 8 key (US gpm is the most common) 
then press the Enter button. 

6.

 

Select the amount of Damping (10 seconds is the most common) then press Enter. Damping is used to 
average the sample values taken. 

7.

 

Select Yes at the Store Measured Data screen then press Enter. 

8.

 

Select No at the Serial Output screen then press Enter. 

9.

 

The storage rate is the frequency at which the meter stores the measured values. It is used for storing 
the  measured data  and can  be  set  for  each  measuring  channel  independently.  Select  how  often you 
wish to have the values stored in memory by using the 2 or 8 key (Every Minute is the most common) 
then press Enter. 

Note: 

If you don't set the storage rate, the default rate or the last rate selected will be used. The storage interval in 

seconds should be at least equal to the number of activated measuring channels. (When 2 measuring channels 

are activated, the storage rate for a channel should be at least 2 seconds.) 

 