CRVW3 3-Channel Vibrating-Wire Datalogger
23
8.1.5 Configuring the CRVW3
Once you have successfully connected to the CRVW3 using
DevConfig
(see
above), you can configure the device. This is an important installation activity
that cannot be overlooked. Key settings must be configured before (or when) a
device is deployed to the field.
Use the
Settings Editor
tab that is available when
DevConfig
is connected to
make configurations to the CRVW3. The following sub-tabs will appear:
Main
,
PakBus
,
Channel 1
,
Channel 2
, and
Channel 3
. CRVW3 devices with
embedded radios will have
Radio
, and
Radio Power Schedule
tabs. The
–RF451 option will also have a
Radio Advanced
tab.
Key settings whose values you should review and set are listed below. As you
place the cursor in the box for each setting, the frame in the lower right corner
of the
DevConfig
screen shows text that describes the setting and how it is used
(online help).
8.1.5.1 Key Settings
[TabName – SettingName]
Main – Station Name
You can give a name to each CRVW3 station and store it in the device. When
planning a network of multiple devices, station naming can assist you in the
organization of your system. (See
Planning your Network
.)
Main – Measurement Interval
This setting specifies how often the CRVW3 should measure the sensors
connected to its active channels. The CRVW3 will repeatedly and
automatically make these measurements at this interval and store the results
until powered down, or re-configured.