CRVW3 3-Channel Vibrating-Wire Datalogger
104
•
Point to MultiPoint Repeater
: Repeater only stations, stand-alone
RF451s
Step 2
Ensure the following settings are identical for every device in the network:
Under
Radio
tab:
•
Frequency Key
•
Network ID
Under
Radio Advanced
tab:
•
Hop Table Version
•
Hop Table Size
•
RF Max Packet Size
•
RF Min Packet Size
•
RF Data Rate
Step 3
For RF451 stand-alone radios, check the
Baud Rate
and
Active Interface
setting and verify that they are correct for your network.
13.2.2 RF407 Network Troubleshooting Summary
Step 1
Using
DevConfig
, ensure the
Radio Enabled
setting is set to
True
for each
CRVW3-RF407 in the network. If the CRVW3-RF407 is to be used as a
repeater, verify the
Is Router
setting is set to
True
(p. 23)
, for information on how to configure the
Is
Router
setting.)
Step 2
Ensure the following settings found on the
Radio
tab are identical for every
device in the network:
•
RF Hop Sequence
•
Network ID
Step 3
For RF407 stand-alone radios, check the
Baud Rate
and
Active Interface
setting and verify that they are correct for your network.