7.4.1 Using an additional CRS45X
One simple way to complete this correction is to deploy an additional CRS45X as a reference
sensor. For this method, in the Select Field Application window, choose Barometer Pressure
Reference (see
Configure CRS45X for measurements
7.4.2 Using a barometer
The barometric pressure data from a nearby Campbell Scientific weather station or other
Campbell Scientific monitoring platform can also be used to compensate for barometric pressure.
The Data tab is used to process and merge the data file from the test sensor with the data file
from the barometric reference sensor. The procedure for doing this follows:
1. Press the browse button next to the Collected Data File field to browse to the data files
retrieved from the CRS45X.
2. Select a data file you wish to be corrected by barometric pressure values.
a. The Field Application Used to Produce Collected Measurement Data information will
be shown based on the collected data file selected.