Section 7. Installation
231
Scan
(100,mSec,100,0)
'Measure Bridge Resistance
BrFull
(Raw_mVperV,1,mV25,1,Vx1,1,2500,True ,True ,0,250,1.0,0)
'Calculate Strain for 1/4 Bridge (1 Active Element)
StrainCalc
(microStrain,1,Raw_mVperV,Zero_mVperV,1,GF_Adj,0)
'Steps (1) & (3): Zero Calibration
'Balance bridge and set Zero_Mode = 1 in numeric monitor. Repeat after
'shunt calibration.
FieldCalStrain
(10,Raw_mVperV,1,0,Zero_mVperV,Zero_Mode,0,1,10,0 ,microStrain)
'Step (2) Shunt Calibration
'After zero calibration, and with bridge balanced (zeroed), set
'KnownRes = to gage resistance (resistance of gage at rest), then set
'Shunt_Mode = 1. When Shunt_Mode increments to 3, position shunt resistor
'and set KnownRes = shunt resistance, then set Shunt_Mode = 4.
FieldCalStrain
(13,MicroStrain,1,GF_Adj,0,Shunt_Mode,KnownRes,1,10,GF_Raw,0)
CallTable
CalHist
NextScan
EndProg
7.7.11.6.3 FieldCalStrain() Quarter-Bridge Shunt Example
With CRBasic example
FieldCalStrain() Calibration
(p. 230)
sent to the CR800, and
the strain gage stable, use the CR1000KD Keyboard/Display or software numeric
monitor to change the value in variable
KnownRes
to the nominal resistance of
the gage,
1000
Ω, as shown in figure
Strain Gage Shunt Calibration Start
(p. 231).
Set
Shunt_Mode
to
1
to start the two-point shunt calibration. When
Shunt_Mode
increments to
3
, the first step is complete.
To complete the calibration, shunt R1 with the 249 kΩ resistor. Set variable
KnownRes
to
249000
. As shown in figure
Strain Gage Shunt Calibration Finish
(p. 232),
set
Shunt_Mode
to
4
. When
Shunt_Mode
=
6
, shunt calibration is
complete.
FIGURE 60: Strain Gage Shunt
Calibration Start
Содержание CR800 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 34: ......
Страница 54: ......
Страница 92: ......
Страница 190: ...Section 7 Installation 190 FIGURE 40 Running Average Frequency Response FIGURE 41 Running Average Signal Attenuation ...
Страница 310: ......
Страница 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Страница 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Страница 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Страница 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Страница 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Страница 456: ......
Страница 462: ......
Страница 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Страница 524: ......
Страница 526: ......
Страница 556: ......
Страница 558: ......
Страница 560: ......
Страница 597: ......