Protect the CR6 from humidity and moisture.
•
Replace the internal lithium battery periodically.
•
Send to Campbell Scientific for factory calibration every three years.
9.1
Protection from Moisture — Details
Protection from Moisture — Overview
(p. 95)
Protection from Moisture — Details
(p. 147)
Protection from Moisture — Products
(p. 670)
When humidity levels reach the dew point, condensation occurs and damage to
CR6 electronics can result. Effective humidity control is the responsibility of the
user. The CR6 module is protected by a packet of silica gel desiccant, which is
installed at the factory. This packet is replaced whenever the CR6 is repaired at
Campbell Scientific. The module should not normally be opened except to replace
the internal lithium battery.
Adequate desiccant should be placed in the instrumentation enclosure to provide
added protection.
9.2
Internal Battery — Details
CAUTION
Fire, explosion, and severe-burn hazard. Misuse or improper
installation of the internal lithium battery can cause severe injury. Do not
recharge, disassemble, heat above 100 °C (212 °F), solder directly to the
cell, incinerate, or expose contents to water. Dispose of spent lithium
batteries properly.
The CR6 contains a lithium battery that operates the clock and SRAM when the
CR6 is not powered. The CR6 does not draw power from the lithium battery while
it is fully powered by a
power supply
(p. 93).
In a CR6 stored at room temperature,
the lithium battery should last approximately three years (less at temperature
extremes). In installations where the CR6 remains powered, the lithium cell
should last much longer.
When a CR6 is put on the shelf for extended periods, the setting
DisableLithium
can be set to disconnect the lithium battery from CR6 circuits, thereby preserving
it for when the CR6 is placed into service. This setting should be used with
caution since disconnecting the lithium battery, either manually or through
settings, will reset the clock and erase SRAM data if primary power is also
Содержание CR6 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 36: ......
Страница 38: ......
Страница 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Страница 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Страница 136: ......
Страница 251: ...Section 7 Installation 251 FIGURE 46 Running Average Frequency Response FIGURE 47 Running Average Signal Attenuation ...
Страница 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Страница 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Страница 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Страница 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Страница 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Страница 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Страница 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Страница 542: ...Section 8 Operation 542 8 11 5 PCCard Memory Card Management FIGURE 128 CR1000KD PCCard Memory Card Management ...
Страница 546: ......
Страница 549: ...Section 9 Maintenance Details 549 FIGURE 133 Separate Back Shell from Module FIGURE 134 Disconnect Battery Connector ...
Страница 552: ......
Страница 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Страница 612: ......
Страница 648: ......
Страница 650: ......
Страница 688: ......
Страница 689: ......