Section 7. Installation
305
Voltage Measurement Instruction Parameters for Dwell Burst
Parameters
Description
Destination
A variable array dimensioned to store all measurements from one input. For
example, the declaration,
Dim
FastTemp(500)
dimensions array
FastTemp()
to store 500 measurements, which is one second of
data at 500 Hz or one-half second of data at 1000 Hz.
The dimension can be any integer from
1
to
65535
.
Count
(was
Repetitions
)
The number of measurements to make on one channel. This number usually
equals the number of elements dimensioned in the
Destination
array.
Valid arguments range from
1
to
65535
.
Voltage Range
The analog-input voltage range to be used during measurements. No change from
standard measurement mode. Use any valid voltage range. However, ranges
appended with
C
cause measurements to be slower.
Single-Ended Channel
The single-ended analog-input terminal number preceded by a dash (
-
). Valid
arguments range from
-
U
1
to
-
U12.
Differential Channel
The differential analog input terminal number preceded by a dash (
-
). Valid
arguments range from
-
U
1
to
-
U11.
Measure Offset
No change from standard measurement mode. For fastest rate, set to
False.
Measurements per Excitation
Must equal the value entered in
Repetitions
.
Reverse Ex
No change from standard measurement mode. For fastest rate, set to
False.
Rev Diff
No change from standard measurement mode. For fastest rate, set to
False.
SettlingTime
Settling time in µs. Range is 100 to 600000 µs. Shorter settling times allow faster
bursts.
[SampleRate (Hz)]
Measurement rate in Hz. Range is 5 to 93750 Hz.Arguments 50 and 60 are valid
for ac rejection but are probably not very useful in burst applications.
Multiplier
No change from standard measurement mode. Enter the proper multiplier. This is
the slope of the linear equation that equates output voltage to the measured
phenomena. Any number greater or less than
0
is valid.
Offset
No change from standard measurement mode. Enter the proper offset. This is the
Y intercept of the linear equation that equates output voltage to the measured
phenomena.
7.7.13.1 Tips — Fast Analog Voltage
•
In the preceding examples, the CR6 disables the auto self-calibration to
reach the stated measurement speeds. Disabling auto self-calibration
increases the risk of measurement errors, especially when the CR6 is
exposed to temperature swings.
Содержание CR6 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 36: ......
Страница 38: ......
Страница 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Страница 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Страница 136: ......
Страница 251: ...Section 7 Installation 251 FIGURE 46 Running Average Frequency Response FIGURE 47 Running Average Signal Attenuation ...
Страница 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Страница 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Страница 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Страница 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Страница 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Страница 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Страница 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Страница 542: ...Section 8 Operation 542 8 11 5 PCCard Memory Card Management FIGURE 128 CR1000KD PCCard Memory Card Management ...
Страница 546: ......
Страница 549: ...Section 9 Maintenance Details 549 FIGURE 133 Separate Back Shell from Module FIGURE 134 Disconnect Battery Connector ...
Страница 552: ......
Страница 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Страница 612: ......
Страница 648: ......
Страница 650: ......
Страница 688: ......
Страница 689: ......