![Campbell CR3000 Micrologger Скачать руководство пользователя страница 517](http://html1.mh-extra.com/html/campbell/cr3000-micrologger/cr3000-micrologger_operators-manual_489090517.webp)
Section 10. Troubleshooting
517
of the host PCs are different. All Campbell Scientific datalogger support
software include an option to change PC PakBus addressing.
See CommMemFree
(p. 578).
10.8.3 Comms Memory Errors
The status array CommsMemFree()
(p. 578,
p. 579, p. 579) may indicate when a
communication memory error occurs. If any of the three CommsMemFree()
array fields are at or near zero, assistance may be required from Campbell
Scientific.
10.9
Troubleshooting — Power Supplies
Related Topics:
• Power Input Terminals — Specifications
• Power Supplies — Quickstart
(p. 39)
• Power Supplies — Overview
(p. 86)
• Power Supplies — Details
(p. 98)
• Power Supplies — Products
(p. 618)
• Power Sources
(p. 99)
• Troubleshooting — Power Supplies
(p. 517)
10.9.1 Troubleshooting Power Supplies — Overview
Power-supply systems may include batteries, charging regulators, and a primary
power source such as solar panels or ac/ac or ac/dc transformers attached to mains
power. All components may need to be checked if the power supply is not
functioning properly.
Check connections. Check polarity of connections.
Base diagnostic: connect the datalogger to a new 12 V battery (a small 12 V
battery carrying a full charge would be a good thing to carry in your
troubleshooting tool kit). Watch the polarity of the connection. + to +, – to –.
If the datalogger powers up and works, troubleshoot the datalogger power supply.
Troubleshooting Power Supplies — Procedures
(p. 518)
includes the following
flowcharts for diagnosing or adjusting power equipment supplied by Campbell
Scientific:
•
Battery-voltage test
•
Charging-circuit test (when using an unregulated solar panel)
•
Charging-circuit test (when using a transformer)
•
Adjusting charging circuit
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......