Section 10. Troubleshooting
507
10.5.3.1 Measurements and NAN
A NAN indicates an invalid measurement.
10.5.3.1.1
Voltage Measurements
The CR3000 has the following user-selectable voltage ranges: ±5000 mV, ±1000
mV, ±200 mV, and ±50 mV. Input signals that exceed these ranges result in an
over-range indicated by a NAN for the measured result. With auto range to
automatically select the best input range, a NAN indicates that either one or both
of the two measurements in the auto-range sequence over ranged. See
Troubleshooting — Auto Self-Calibration Errors.
A voltage input not connected to a sensor is floating and the resulting measured
voltage often remains near the voltage of the previous measurement. Floating
measurements tend to wander in time, and can mimic a valid measurement. The
C (open input detect/common-mode null) range-code option is used to force a
NAN result for open (floating) inputs.
10.5.3.1.2
SDI-12 Measurements
NAN is loaded into the first SDI12Recorder() variable under the following
conditions:
•
CR3000 is busy with terminal commands
•
When the command is an invalid command.
•
When the sensor aborts with CR LF and there is no data.
•
When 0 is returned for the number of values in response to the M! or C!
command.
10.5.3.2 Floating-Point Math, NAN, and ±INF
Related Topics:
• Floating-Point Arithmetic
(p. 170)
• Floating-Point Math, NAN, and ±INF
(p. 507)
• TABLE: Data Types in Variable Memory
(p. 137)
Table Math Expressions and CRBasic Results
(p. 508)
lists math expressions, their
CRBasic form, and IEEE floating point-math result loaded into variables declared
as FLOAT or STRING.
10.5.3.3 Data Types, NAN, and ±INF
NAN and ±INF are presented differently depending on the declared-variable data
type. Further, they are recorded differently depending on the final-memory data
type chosen compounded with the declared-variable data type used as the source
(TABLE: Variable and FS Data Types with NAN and ±INF
(p. 508)
). For example,
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......