CR10TCR THERMOCOUPLE REFERENCE
3
TABLE 3. Temperature, Resistance, and
Datalogger Output
Temperature
Resistance
Output
°
C
OHMS
°
C
-40.00
4067212
-39.18
-38.00
3543286
-37.55
-36.00
3092416
-35.83
-34.00
2703671
-34.02
-32.00
2367900
-32.13
-30.00
2077394
-30.18
-28.00
1825568
-28.19
-26.00
1606911
-26.15
-24.00
1416745
-24.11
-22.00
1251079
-22.05
-20.00
1106485
-20.00
-18.00
980100
-17.97
-16.00
869458
-15.95
-14.00
772463
-13.96
-12.00
687276
-11.97
-10.00
612366
-10.00
-8.00
546376
-8.02
-6.00
488178
-6.05
-4.00
436773
-4.06
-2.00
391294
-2.07
0.00
351017
-0.06
2.00
315288
1.96
4.00
283558
3.99
6.00
255337
6.02
8.00
230210
8.04
10.00
207807
10.06
12.00
187803
12.07
14.00
169924
14.06
16.00
153923
16.05
18.00
139588
18.02
20.00
126729
19.99
22.00
115179
21.97
24.00
104796
23.95
26.00
95449
25.94
28.00
87026
27.93
30.00
79428
29.95
32.00
72567
31.97
34.00
66365
33.99
36.00
60752
36.02
38.00
55668
38.05
40.00
51058
40.07
42.00
46873
42.07
44.00
43071
44.05
46.00
39613
46.00
48.00
36465
47.91
50.00
33598
49.77
52.00
30983
51.59
54.00
28595
53.35
56.00
26413
55.05
58.00
24419
56.70
60.00
22593
58.28
TABLE 4. Polynomial Coefficients
Coefficient
Value
C0
-53.4601
C1
90.807
C2
-83.257
C3
52.283
C4
-16.723
C5
2.211
6. ELECTRICALLY NOISY
ENVIRONMENTS
AC power lines can be the source of electrical
noise. If the datalogger is in an electronically
noisy environment, the CR10TCR temperature
measurement should be measured with the AC
half bridge (Instruction 5) with the 60 Hz rejection
integration option on the CR10 (see Section 13
of the datalogger manual for more information
on noise). Instruction 11's fast integration will
not reject 60 Hz noise.
Example 2. Sample CR10 Instructions Using
AC Half Bridge
01:
P5
AC Half Bridge
01:
1
Rep
02:
22
7.5 mV 60 Hz rejection Range
03:
1
IN Chan
04:
3
Excite all reps w/EXchan 1
05: 2000
mV Excitation
06:
15*
Loc [:Ref_Temp]
07:
800
Mult
08:
0
Offset
02:
P55
Polynomial
01:
1
Rep
02:
1*
X Loc Air Temp
03:
15*
F(X) Loc [:Ref_Temp]
04:
-53.46
C0
05:
90.807
C1
06:
-83.257
C2
07:
52.283
C3
08:
-16.723
C4
09:
2.211
C5
*
Proper entries will vary with program and
input location assignments.