H-1
Appendix H. Output Variables
The CPEC300 program stores data in several output tables. Details are given
for each table.
ts_data
The primary output table is
ts_data
table which gives time-series data. This
table stores each sample of the raw CPEC300 data (ten records per sec). The
CPEC300 program stores this table in multiple files on the memory card, with
a new file started each day at midnight. The size of these daily files depends on
the compile flags. With the default settings, each file is approximately 62 kB
and a 2 GB memory card will store data for approximately four weeks before
the CPEC300 program begins to delete old files to make room for new files. If
no valve module is used, the time-series files are slightly smaller,
approximately 55 kB. The largest files will occur when the compile switch is
set to store all diagnostic information, and use the valve module and scrub
modules. These daily files are approximately 120 kB in size. See Section
(p. 36)
, for details on setting program constants. If more storage is
required, a 16 GB memory card, the CFMC16G, is available.
The values stored in table
ts_data
are shown in TABLE
. Values that are
stored or not stored depending on compile switch settings have a V, D or S in
the first column of TABLE
•
V means the value is stored only if
VALVE_MODULE
=
True
•
D means the value is stored only if SaveAll
_diagnostics
=
True
•
S means the value is stored only if
SCRUB_MODULE
=
True
If multiple compile switches are listed for a value, the value is stored in the
table only if all of the compile switches are
True
. If no compile switches are
listed for a value, the value is always stored in the table.
TABLE H-1. Values stored in table ts_data
Compile
Switch
Variable
Name
Units
Description
TIMESTAMP
TS
Date and time the record was measured
RECORD
RN
Record number
EC_mode
Boolean flag: True if the pump is on and no zero/span sequence
is running
diag_cpec
CPEC300 diagnostic word
Ux
m∙s
-1
Wind velocity X component
Uy
m∙s
-1
Wind velocity Y component
Uz
m∙s
-1
Wind velocity Z component
Ts
ºC
Sonic virtual temperature
diag_sonic
CSAT3A diagnostic word
CO2
µmol∙mol
-1
CO
2
mixing ratio, relative to dry air
H2O
mmol∙mol
-1
H
2
O vapor mixing ratio, relative to dry air
diag_irga
EC155 diagnostic word
cell_tmpr
ºC
Temperature of the sample cell
cell_press
kPa
Pressure in the sample cell
diff_press
kPa
Differential pressure (sample cell, relative to ambient)
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...