D-1
Appendix D. System Diagnostic Word
The system diagnostic word,
system_diag,
is an aggregate or sum of the system
diagnostic bits or flags listed below in TABLE
TABLE D-1. Description of Bits for System Diagnostic Word.
Bit
Decimal
Name
Function
0
1
sonc_er
Sonic error. If any error condition or diagnostic flag is set on the
sonic anemometer, this bit is set.
1
2
irga_er
IRGA error. If any error condition or diagnostic flag is set on the
gas analyzer, this bit is set.
2
4
pump_tmpr_er
Pump temperature error. If pump temperatre is below 0 deg C or
above 55 deg C, this bit is set.
3
8
pump_flow_er
Pump flow error. If pump flow is more than ±0.5 L
∙
min
-1
from the
pump flow set point (default set point is 8 L
∙
min
-1
), this bit is set.
4
16
valv_tmpr_er
Valve temperature error. If valve module temperature is below 0
deg C or above 60 deg C, this bit is set.
5
32
valv_flow_er
Valve flow error. If zero or span gas flow through valve module is
more than ±0.5 L
∙
min
-1
from the valve flow set point (default set
point is 1 L
∙
min
-1
), this bit is set.
6
64
scrb_tmpr_er
Scrub temperature error. If scrub temperature is below 5 deg C or
above 50 deg C, this bit is set.
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...