CFM100 CompactFlash

®

 Module 

14 

7.3.2.2  Reinserting the Card 

If the same card is inserted again into the CFM100, the datalogger will store all 
data to the card that has been generated since the card was removed that is still 
in the CPU memory. If the data tables have been left on the card, new data will 
be appended to the end of the old files. If the data tables have been deleted, 
new ones will be generated.  

Check the status of the card before leaving the datalogger. If a CF 
card was not properly accepted, the CFM100 will flash orange. In 
that case, the CF card needs to be reformatted and all data erased. 
Formatting or erasing a CF card can  be done on a PC or 
datalogger. The procedure for formatting a CF card is explained 
in Appendix 

A

CF Card Maintenance 

(p. A-1)

. 

7.3.2.3  Card Swapping 

When transporting a CF card to a computer to retrieve data, most users will 
want to use a second card to ensure that no data is lost. For this method of 
collection, use the following steps. 

1.

 

Insert formatted card (“CF-A”) in CFM100 attached to datalogger. 

2.

 

Send program containing CardOut() instruction(s). 

3.

 

When ready to retrieve data, press the Initiate Removal button to remove 
card. The LED will show red while the most current data is stored to the 
card and then will turn green. Eject card while LED is green. 

4.

 

Put in clean card (“CF-B”).  

5.

 

Use CardConvert to copy data from CF-A to PC and convert. The default 
CardConvert filename will be TOA5_stationname_tablename.dat. Once 
the data is copied, use Windows Explorer to delete all data files from the 
card
. NOTE: Windows98 and WindowsME users need to shift-delete to 
completely delete files. Using standard delete may create an invisible 
recycle bin on the CF card.  

6.

 

At the next card swap, eject CF-B and insert the clean CF-A. 

7.

 

Running CardConvert on CF-B will result in separate data files containing 
records since CF-A was ejected. CardConvert can increment the filename 
to TOA5_stationname_tablename_0.dat. 

8.

 

The data files can be joined using a software utility such as WordPad or 
Excel.  

CardConvert File 

CF-A Record Numbers  CF-B Record Numbers 

TOA5_tablename.dat 

0-100 

 

TOA5_tablename.dat 

 

101-1234 

TOA5_tablename.dat 

1235-…. 

 

 

NOTE 

Содержание CompactFlash CFM100

Страница 1: ...INSTRUCTION MANUAL CFM100 CompactFlash Module Revision 9 16 C o p y r i g h t 2 0 0 0 2 0 1 6 C a m p b e l l S c i e n t i f i c I n c...

Страница 2: ......

Страница 3: ......

Страница 4: ...er Campbell Scientific s shipping address is CAMPBELL SCIENTIFIC CANADA CORP RMA _____ 14532 131 Avenue NW Edmonton Alberta T5L 4X4 Canada For all returns the client must fill out a Statement of Produ...

Страница 5: ...protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit climbing by other persons Take reasonabl...

Страница 6: ...ounce 28 35 g 1 lb pound weight 0 454 kg Pressure 1 psi lb in2 68 95 mb Volume 1 US gallon 3 785 litres In addition part ordering numbers may vary For example the CABLE5CBL is a CSI part number and k...

Страница 7: ......

Страница 8: ...Data Retention 5 6 Specifications 5 6 1 Power 5 7 Operation 6 7 1 File Formats 6 7 1 1 Data Files 6 7 1 2 Program Files 6 7 1 3 Power Up Files Powerup ini 6 7 1 3 1 Creating and Editing Powerup ini 7...

Страница 9: ...2 A 2 Checking CF Card Integrity A 3 Figures 5 1 CompactFlash Module 4 7 1 CardConvert 13 Tables 7 1 Powerup ini Commands 8 CRBasic Examples 7 1 Powerup ini Code 8 7 2 Run Program on Power up 9 7 3 Fo...

Страница 10: ...manual The CFM100 is rugged but it should be handled as a precision scientific instrument There are no user serviceable parts inside the module The 28033 surge suppressor and or a shielded 10base t E...

Страница 11: ...ogramming the Datalogger to Send Data to the CFM100 The CardOut instruction is used in the datalogger program to send data to the CF card The CardOut instruction must be entered within each DataTable...

Страница 12: ...trieval is discussed in detail in Section 7 3 Data Retrieval p 12 Removing a card while it is active can cause garbled data and can actually damage the card Always press the Initiate Removal button an...

Страница 13: ...id green when it is OK to remove the card solid orange to indicate an error and flashing orange if the card has been removed and has been out long enough that CPU memory has wrapped and data is being...

Страница 14: ...x 2 6 in Weight 132 5 g Operating Temperature Range 35 to 65 C 55 to 85 C optional Typical Access Speed 200 to 400 kbits s 1 Memory Configuration User selectable for either ring style default or fill...

Страница 15: ...some additional memory allocated When the file is converted this will result in additional records if no lapses occurred If more lapses occur than were anticipated there may be fewer records in the f...

Страница 16: ...move the CF card or delete the powerup ini file after the powerup ini upload 7 1 3 1 Creating and Editing Powerup ini Powerup ini is created with a text editor and saved as powerup ini Some text edito...

Страница 17: ...ce Command 1 2 and 13 Run On Power up If the copy first application above succeeds the new Run On Power up program is accepted If the copy fails no change will be made to the Run On Power up program C...

Страница 18: ...RD run always erase CRD data files 13 toobigforcpu cr1 crd CRBasic Example 7 6 Run Program Always Erase CF Data Run always erase CRD data files 13 pwrup_1 cr1 crd CRBasic Example 7 7 Run Program Now E...

Страница 19: ...bles with a negative value for size The partitioning algorithm attempts to have the tables full at the same time 7 2 2 Program Examples 7 2 2 1 Ring Mode The following program outputs the maximum and...

Страница 20: ...he card once a second The first parameter of the CardOut instructions is 1 which sets the tables on the card to fill and stop mode The second parameter is 1000 so 1000 records will be written to each...

Страница 21: ...be retrieved through a communication link to the datalogger or by removing the card and carrying it to a computer 7 3 1 Via a Communication Link Data can be transferred to a computer via a communicati...

Страница 22: ...wever it will be necessary to convert the file format before using the data When dealing with large data files it may be faster to use an external card reader such as pn 17752 rather than a PC card sl...

Страница 23: ...sert formatted card CF A in CFM100 attached to datalogger 2 Send program containing CardOut instruction s 3 When ready to retrieve data press the Initiate Removal button to remove card The LED will sh...

Страница 24: ...ontrol A 1 1 Windows Explorer To format card using Windows Explorer 1 Insert CF card into CF adapter or CF reader 2 Windows Explorer should identify a drive as a removable disk F 3 Select that drive a...

Страница 25: ...e CR1000KD 1 Insert CF card into CFM100 2 From the main menu of CR1000KD choose PCCard 3 Choose Format Card 4 Press Yes to proceed A 1 3 LoggerNet File Control To format card using LoggerNet File Cont...

Страница 26: ...ess Format 6 Press Yes to confirm A 2 Checking CF Card Integrity The Windows Check Disk tool can be used to check the integrity of a CF card To access the Check Disk tool 1 Insert CF card into CF read...

Страница 27: ...Appendix A CF Card Maintenance A 4 5 Navigate to the Tools tab 6 Press Check 7 Select Scan and Repair Drive...

Страница 28: ......

Страница 29: ...ua Apinag s nbr 2018 Perdizes CEP 01258 00 S o Paulo SP BRASIL www campbellsci com br vendas campbellsci com br Campbell Scientific Canada Corp 14532 131 Avenue NW Edmonton AB T5L 4X4 CANADA www campb...

Отзывы: