Campbell CMP3 Скачать руководство пользователя страница 22

CMP3 Pyranometer 

12 

7.4.2  Multiplier 

The multiplier converts the millivolt reading to engineering units. The 

calibration supplied by the manufacturer gives the output of the sensor (c) as 

microvolts per W/m^2. As the data logger voltage measurement instructions 

give a default output in millivolts, the following equation should be used to 

calculate the multiplier (m) to give the readings in W/m

2

 

m = 1000/c 

Other units can be used by adjusting the multiplier as shown in TABLE 

7-2

. 

TABLE 7-2.  Multipliers Required for Flux Density and Total Fluxes 

Units 

Multipliers 

Output Processing 

W/m

2

 

Average 

MJ/m

2

 

m × t × 0.000001 

Total 

kJ/m

–2

 

m × 

× 0.001 

Total 

cal/cm

2

 

m × t × 0.0239 × 0.001 

Total 

cal/cm

2

/min 

m × 1.434 × 0.001 

Average 

m = calibration factor in W/m

2

/mV 

t = data logger program execution interval in seconds 

 

7.4.3  Offset 

The offset will normally be fixed at zero as the sensor should output no 

significant signal in dark conditions. In practice, because of the nature of 

thermopile detector sensors, there will be some offset in dark conditions; 

sometimes this offset can give negative light readings. This offset varies with 

several factors, such as rate of change of sensor temperature, so it cannot be 

removed with a fixed offset. Some users may wish to remove small negative 

readings by including code after the measurement instructions that sets 

negative readings to zero. 

7.4.4  Output Format Considerations  

When using the Campbell Scientific floating point data format to store data, the 

largest number the data logger can store in final storage is 6999 in low 

resolution mode (FP2) and 99999 in high resolution mode (if available). If the 

measurement value is totalized, there is some danger of over-ranging the output 

limits. To avoid this issue, store the data in the in IEEE4 format, which can 

represent a wider range of numbers. 

8.  Maintenance and Troubleshooting 

All factory repairs and recalibrations require a returned material 

authorization (RMA) and completion of the “Declaration of 

Hazardous Material and Decontamination” form.  Refer to the 

Assistance

  page at the beginning of this manual for more 

information. 

NOTE 

Содержание CMP3

Страница 1: ...CMP3 Pyranometer Revision 02 2020 Copyright 2006 2020 Campbell Scientific CSL I D 332...

Страница 2: ......

Страница 3: ...under guarantee or not Please state the faults as clearly as possible and if the product is out of the guarantee period it should be accompanied by a purchase order Quotations for repairs can be give...

Страница 4: ......

Страница 5: ...digital cell phones and aerials may also not be applicable according to your locality Some brackets shields and enclosure options including wiring are not sold as standard items in the European marke...

Страница 6: ......

Страница 7: ...beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit cl...

Страница 8: ......

Страница 9: ...6 Specifications 5 7 Installation 6 7 1 Siting 6 7 2 Mounting to an Instrument Mount 7 7 2 1 Required Tools 7 7 2 2 Mounting Procedure 7 7 3 Wiring to the Data Logger 9 7 4 Programming 10 7 4 1 Input...

Страница 10: ...s 7 1 Pyranometer installation 6 7 2 CMP3 Schematic 10 Tables 7 1 Wire Colour Function and Data Logger Connection 10 7 2 Multipliers Required for Flux Density and Total Fluxes 12 CRBasic Example B 1 C...

Страница 11: ...l Inspection Check the contents of the shipment If there is a shortage see Section 3 1 Ships With p 1 contact Campbell Scientific If any damage has occurred during transport immediately file a claim w...

Страница 12: ...It is included in installations of LoggerNet RTDAQ PC400 or PC200W The following procedure also describes programming with Short Cut 1 Open Short Cut and click Create New Program 2 Double click the da...

Страница 13: ...b to see how the sensor is to be wired to the data logger Click OK after wiring the sensor 5 Repeat steps three and four for other sensors Click Next 6 In Output Setup type the scan rate meaningful ta...

Страница 14: ...l Scientific Cable length is user specified The CMP3 pyranometer is designed for continuous outdoor use Due to its flat spectral sensitivity from 300 to 2800 nm it can be used in natural sunlight unde...

Страница 15: ...l Scientific CRBasic data loggers CR6 CR3000 CR1000X CR800 series CR300 series CR1000 6 Specifications ISO classification Second Class Response time 95 18 s Zero offset due to 200 W m2 thermal radiati...

Страница 16: ...r you See Section 4 QuickStart p 2 for a Short Cut tutorial 7 1 Siting The CMP3 is usually installed horizontally but can also be installed at any angle including an inverted position In all cases it...

Страница 17: ...hemisphere point the cable toward the North Pole 7 2 Mounting to an Instrument Mount 7 2 1 Required Tools Tools required for installation on a tripod or tower Small and medium Phillips screwdrivers 8...

Страница 18: ...by tightening the U bolt nuts 3 Loosely mount the pyranometer on the mounting stand Do not fully tighten the two mounting screws CM225 Stand CM200 Series Crossarm U bolt Nuts CM225 mounting holes use...

Страница 19: ...still correctly levelled and adjust as necessary 6 Attach the white plastic sun shield to the pyranometer 7 Route the sensor cable along the underside of the crossarm to the tripod tower and to the in...

Страница 20: ...onnection Terminal Single Ended Data Logger Connection Terminal White Signal High U configured for differential input1 DIFF H differential high analogue voltage input U configured for single ended ana...

Страница 21: ...ed measurement compare the results of single ended and differential measurements made under the same conditions Possible sources of electrical noise include power lines electric pumps or motors If the...

Страница 22: ...e of thermopile detector sensors there will be some offset in dark conditions sometimes this offset can give negative light readings This offset varies with several factors such as rate of change of s...

Страница 23: ...ientific recommends recalibrating every two years The sensor should be returned to Campbell Scientific the manufacturer or a calibration lab with facilities to calibrate radiation sensors 8 2 Troubles...

Страница 24: ...yranometer 14 1 Check for the presence of strong sources of electromagnetic radiation radar radio 2 Check the condition and the connection of the sensor shield wire 3 Check the condition of the sensor...

Страница 25: ...can now be edited for additional refinement Once the file is edited with CRBasic Editor Short Cut can no longer be used to edit the program it created 2 To add the Short Cut wiring information into t...

Страница 26: ......

Страница 27: ...e B 1 CR1000X Example Program CR1000X Declare Variables and Units Public Solar_Wm2 Public Solar_MJ Units Solar_Wm2 W m Units Solar_MJ MJ m Hourly Data Table DataTable Hourly True 1 DataInterval 0 60 M...

Страница 28: ...If Solar_Wm2 0 Then Solar_Wm2 0 Calculate units in MJ where MJ m t 0 000001 m Solar_Wm2 from above and t 10 scan interval Solar_MJ Solar_Wm2 0 00001 Call Data Tables and Store Data CallTable Hourly Ca...

Страница 29: ......

Страница 30: ...cennes France Phone 0033 0 1 56 45 15 20 Email info campbellsci fr Website www campbellsci fr Germany Location Bremen Germany Phone 49 0 421 460974 0 Email info campbellsci de Website www campbellsci...

Отзывы: