CMP6-L, CMP11-L, and CMP21-L Pyranometers
6. Specifications
6.1 Pyranometers
TABLE 6-1. CMP-series Specifications
Specification CMP6
CMP11 CMP21
ISO Classification
First Class
Secondary Standard
Maximum irradiance
2000 W
•
m
–2
4000
W
•
m
–2
Spectral range
(50% points)
285 to 2800 nm
Response time (95 %)
<18 s
<5 s
Expected daily
uncertainty
<5% <2%
Zero offset due to thermal
radiation
(200 W
•
m
–2
)
<15 W
•
m
–2
<7
W
•
m
–2
Zero offset due to
temperature change
(5 K
•
hr
–1
)
<4 W
•
m
–2
<2
W
•
m
–2
Non-stability
(change/year)
<1 %
<0.5%
Non-linearity
(0 to 1000 W
•
m
–2
)
<1% <0.2%
Directional error
(up to 80° with 1000
W
•
m
–2
beam)
<20 W
•
m
–2
<10
W
•
m
–2
Tilt error
(at 1000 W
•
m
–2
)
<1% <0.2%
Level accuracy
0.1°
Operating temperature
–40° to 80°C
Temperature dependence
of sensitivity
<4% (–10° to +40°C)
<1% (–10° to +40°C)
<1% (–20° to +50°C)
Sensitivity
5 to 20 µV / W
•
m
–2
7 to 14 µV / W
•
m
–2
Typical signal output for
atmospheric applications
0 to 20 mV
0 to 15 mV
Weight
0.6 kg (1.3 lb) without cable;
0.9 kg (2 lb) with 10 m (33 ft) cable
Impedance* 20
to
200
Ω
10
to
100
Ω
* Impedance is defined as the total electrical impedance at the radiometer output connector fitted to the housing. It