AM25T Solid-State Thermocouple Multiplexer
9
recommends use of Teflon®, polyethylene, or polypropylene insulation around
individual conductors. Do not use PVC as conductor insulation. PVC may be
used as a cable jacket.
With long lead lengths, a delay within the measurement instruction will allow
the capacitance of the lead wires to discharge before the measurement is made.
Consult the measurement section of the datalogger manual for more
information.
8.2 Datalogger Programming
Short Cut
can build many program configurations for various supported
sensors providing a quick way to generate a program and wiring diagram
(FIGURE
8-1
).
Short Cut
can be downloaded free of charge
(
www.campbellsci.com
).
FIGURE 8-1. Short Cut
Short Cut
is the best source for up-to-date datalogger programming code.
Programming code is needed when:
•
Creating a program for a new datalogger installation
•
Adding sensors to an existing datalogger program
If the data acquisition requirements are simple, a datalogger program can be
created and maintained exclusively with
Short Cut.
If the data acquisition
needs are more complex, the files that
Short Cut
creates are a great source for
programming code to start a new program or add to an existing custom
program.