![Campbell AM16 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/campbell/am16/am16_instruction-manual_3762330026.webp)
AM16/32A Relay Analog Multiplexer
'CR5000 Example Program to measure 16 100 ohm Platinum Resistance Thermometers ‘connected to an
AM16/32A multiplexer used in the 4x16 configuration. The program also ‘measures 6 copper constantan
'thermocouples.
'The Thermocouples are connected to differential channels 1-6.
'Declare Variables:
Public TRef, TCTemp(6), PRTResist(16), PRTTemp(16)
Dim I
'Counter for setting Array element to correct value for mux measurement
'Declare Output Table for 15 minute averages:
DataTable (Avg15Min,1,-1)
DataInterval
(0,5,Min,10)
Average
(1,TRef,IEEE4,0)
Average
(6,TCTemp(),IEEE4,0)
Average
(16,PRTTemp(),IEEE4,0)
EndTable
BeginProg
Scan
(60,Sec,3,0)
PanelTemp
(TRef,250)
TCDiff (TCTemp(),6,mV20C ,1,TypeT,TRef,True ,0,250,1.0,0)
Portset
(1
,1)
'Set C1 high to Enable Multiplexer
Delay
(0,150,mSec)
I=0
SubScan(0,sec,16)
'Pulse C2 (Set High, Delay, Set Low) to clock multiplexer
Portset
(2,1
)
Delay (0,20,mSec)
Portset
(2,0)
I=I+1
'The Resistance measurement measures the PRT resistance:
Resistance
(PRTResist(I),1,mV50,7,Ix1,1,500,True
,True
,0,250,0.01,0)
'With a multiplier of 0.01 (1/100) the value returned is R/Ro (Resist/Resist @ 0 deg)
'the required input for the PRT temperature calculation instruction.
NextSubScan
Portset
(1
,0)
'Set C1 Low to disable Multiplexer
‘Calculate the Temperature from R/Ro:
PRT
(PRTTemp(1),16,PRTResist(1),1.0,0)
CallTable
Avg15Min
‘Call the DataTable
NextScan
EndProg
5.3.2 CR1000, CR800, and CR850 Programming
Although the following example is a CR1000 program, a similar
program can be used for the CR800 or CR850. This CR1000
program uses the AM16/32A to measure 48 CS616 probes connected
in the 4x16 configuration. The program also measures datalogger
battery voltage and temperature.
20
Содержание AM16
Страница 6: ...This is a blank page...
Страница 44: ...This is a blank page...
Страница 45: ...This is a blank page...