Model 109B & 109 BAM Temperature Probe 

Steinhart & Hart  - Tabulated values

-0.005

0

0.005

0.01

0.015

0.02

0.025

0.03

-50

-40

-30

-20

-10

0

10

20

30

40

50

60

70

Temperature Degrees C

Error Degrees C

 

FIGURE 2-1.  Steinhart and Hart 

Worst Case Errors in 109 Temperature Measurement

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

-50

-40

-30

-20

-10

0

10

20

30

40

50

60

70

Temperature Degrees C

Po

ssib

le Erro

r Deg

rees C

Thermistor Tolerance

CR200 Bridge Measurement Error (0.06% of 2.4 mV)

CR10X Bridge Measurement Error, CR200 Resolution (0.5 mV, 0.6 mV)

24.9 k Ohm Fixed Resistor Tolerance (0.1% + 10ppm/degC away from 25deg)

 

FIGURE 2-2.  Possible Errors 

3.  Installation 

3.1  Soil Temperature 

The 109B is suitable for burial.  Whenever possible it should be placed 
horizontally at the desired depth to avoid thermal conduction from the surface 
to the thermistor.  The recommended maximum burial depth is equivalent to 86 
psi. 

Содержание 109B

Страница 1: ...INSTRUCTION MANUAL 109B 109BAM Temperature Probe April 2010 Copyright 2010 Campbell Scientific Canada Corp...

Страница 2: ...glect accidents of nature or shipping damage This warranty is in lieu of all other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CSC is no...

Страница 3: ...CRBasic Examples 6 5 1 1 1 Sample Program for CR200 X Series Datalogger 7 5 1 1 2 Sample Program for CR1000 Datalogger 7 5 1 1 3 Sample Program for CR1000 Datalogger AM16 32B 4x16 mode 8 5 2 Edlog 8 5...

Страница 4: ...each single ended channel used This will vary depending on the multiplexer configuration and the number of sensors measured The model name 109B will be used in this manual to refer to both 109B and 10...

Страница 5: ...tion in air It is rated as slow burning when tested according to U L 94 H B and will pass FMVSS302 Local fire codes may preclude its use inside buildings NOTE 2 Accuracy The overall probe accuracy is...

Страница 6: ...rature Degrees C Possible Error Degrees C Thermistor Tolerance CR200 Bridge Measurement Error 0 06 of reading 2 4 mV CR10X Bridge Measurement Error CR200 Resolution 0 5 mV 0 6 mV 24 9 k Ohm Fixed Resi...

Страница 7: ...physically limited by the number of lead wires that can be inserted into a single voltage excitation terminal approximately six TABLE 4 1 109B Connections to Campbell Scientific Dataloggers Colour Hea...

Страница 8: ...COM ODD H Vx 1 E 1 Vx 1 Signal Return 1 White COM ODD L SE 1 SE 1 SE 1 Signal Return 2 Black COM EVEN H SE 2 SE 2 SE 2 Signal Return 3 Green COM EVEN L SE 3 SE 3 SE 3 Shield Clear COM G The wiring ass...

Страница 9: ...gths prior to programming your datalogger Measurement details are provided in Section 6 5 1 CRBasic In the CR200 X series CR800 CR850 CR1000 and CR3000 dataloggers Instruction Therm109 is used to meas...

Страница 10: ...terval 0 10 min Average 1 Air_Temp 0 EndTable BeginProg Scan 1 sec Measure the temperature Therm109 Air_Temp 1 1 Ex1 1 0 0 Call the data table CallTable AvgTemp NextScan EndProg 5 1 1 2 Sample Program...

Страница 11: ...9BAM_C 1 3 1 1 0 _60Hz 1 0 Switch to next AM16 32 Multiplexer channel set PulsePort 1 10000 109BAM Temperature Probe 3 wire measurements T109BAM_C 4 5 6 on the AM16 32B Therm109 T109BAM_C 4 3 1 1 0 _6...

Страница 12: ...ution Interval seconds 1 AC Half Bridge P5 1 1 Reps 2 25 2500 mV 60 Hz Rejection Range 3 1 SE Channel 4 1 Excite all reps w Exchan 1 5 2500 mV Excitation 6 1 Loc V_Vx 7 1 0 Mult 8 0 0 Offset 2 Z 1 X P...

Страница 13: ...2 9 Loc Air_Temp Table 2 Program 02 0 0000 Execution Interval seconds Table 3 Subroutines End Program 5 3 Electrically Noisy Environments AC power lines pumps and motors can be the source of electric...

Страница 14: ...w P5 to convert the measurement result to temperature are still required 1 Excite Delay SE P4 1 1 Reps 2 25 2500 mV 60 Hz Rejection Range Delay must be zero 3 1 SE Channel 4 1 Excite all reps w Exchan...

Страница 15: ...trusion For all factory repairs and recalibrations customers must get a returned material authorization RMA Customers must also properly fill out a Declaration of Hazardous Material and Decontaminatio...

Страница 16: ...5 4 Make sure the clear shield wire is connected to datalogger ground and the datalogger is properly grounded For all factory repairs customers must get an RMA Customers must also properly fill out a...