MODEL 105T THERMOCOUPLE PROBE
INSTRUCTION MANUAL
PRELIMINARY: 31/01/99
COPYRIGHT (c) 1999 CAMPBELL SCIENTIFIC, INC.
Страница 1: ...MODEL 105T THERMOCOUPLE PROBE INSTRUCTION MANUAL PRELIMINARY 31 01 99 COPYRIGHT c 1999 CAMPBELL SCIENTIFIC INC ...
Страница 2: ......
Страница 3: ...t accidents of nature or shipping damage This warranty is in lieu of all other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CSC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorization To obtain a Return Merchandise Authorization RMA contact CAMPBELL SCIENTIFIC CANA...
Страница 4: ......
Страница 5: ...d 2 SPECIFICATIONS Probe ANSI Thermocouple Type T Temperature Range up to 50 C Seebeck Coef 38 µV C 0 C Interchangeability 73 C 0 365 C 20 C 0 1 C 50 C 0 25 C Cable Conductors 24 AWG Blue Copper Red Constantan Clear Shield Insulation Extruded TPR Jacket Extruded TPR Diameter 0 2 inches 3 INSTALLATION The 105T is designed for burial in soils or submersion in water up to 50 feet 4 WIRING Connection ...
Страница 6: ...is a thermistor already located in the wiring panel of the CR23X 21X and CR7 for this purpose For the CR10 X the wiring panel temperature must be measured using a Thermocouple Reference Thermistor CR10XTCR or CR10TCR When installed on the CR10 X the reference thermistor should lie between the two analog input terminal strips Example 1 program illustrates programming for the CR10 X using instructio...
Страница 7: ... 1 0 Mult 6 0 0 Offset Knowing the wiring panel temperature we now measure the 105T voltage and use the type T equation to calculate the 105T temperature A multiplier of 1 and offset of 0 yield the temperature in C in input location 2 TC_Temp 2 Thermocouple Temp DIFF P14 1 1 Reps 2 21 2 5 mV 60 Hz Rejection Range 3 2 DIFF Channel 4 1 Type T Copper Constantan 5 1 Ref Temp Deg C Loc Ref_Temp 6 2 Loc...
Страница 8: ...equation to calculate the 105T temperature A multiplier of 1 and offset of 0 yield the temperature in C in input location 2 TC_Temp 2 Thermocouple Temp DIFF P14 1 1 Reps 2 21 10 mV 60 Hz Reject Slow Range 3 1 DIFF Channel 4 1 Type T Copper Constantan 5 1 Ref Temp Deg C Loc Ref_Temp 6 2 Loc TC_Temp 7 1 0 Mult 8 0 0 Offset ...