PM 1000288 000 00
Device handbook SIRAX BT5100
5/32
3. Device overview
The SIRAX BT5100 is a top-hat rail mounted Transducer and used to measure and convert AC Voltage input into an proportional DC current or
voltage output signal. Input Voltage and Output Voltage/Current is displayed on LCD and indicated by LED’s.
4. Mechanical mounting
The SIRAX BT5100 is designed for panel mounting.
4.1 Mounting
Dimensional drawing BT5100: See section 10
Any mounting position is possible. Device may be clipped onto a top-hat rail according EN50022.
4.2 Demounting of the device
Disassembly of the device requires that all connected wires be without current. First, remove all push terminals and the wires of the current and
voltage inputs. Ensure that possible current transformers are short-circuited before the current connections on the device are opened. Release
the transducer from a top-hat rail.
Please ensure that the operating temperature limits are not exceeded when determining
the place of mounting (place of measurement):
0 … +45 °C