To reassemble a round-body meter, follow the steps below:
Insert a bearing support in the meter body, ensuring that the flow arrow
is pointing in the direction of flow and the punch marks on the support
and meter body are aligned (Figure 7, page 22).
Install the three socket-head capscrews to secure the bearing support,
and tighten them with a standard Allen wrench.
Insert the rotor into the opposite end of the meter body, ensuring that
the flow arrow is pointing in the direction of flow. Handle the rotor with
care to avoid chipping the rotor shaft.
Insert the the second bearing support into the meter body, aligning
punch marks as described in step 1.
Install the three socket-head capscrews to secure the bearing support,
and tighten them with a standard Allen wrench.
Spin the rotor by hand. If the rotor does not turn feely or stops abruptly,
disassemble the meter and check the rotor for damage.
1.
2.
3.
4.
5.
6.