background image

 

5

.1 Testing

 Menu Definitions

:

      

T

1

   Version                  Displays Software Version.  

      

T

2   

Display           

 Lights all display segments and indicating LED s

      

T

3   

Buttons                Press the Zero key, b1 will be displayed.

     Press the GRS/NET key, b2 will be displayed.

     

Press the Tare key, b3 will be displayed.
Press the lb/kg key, b4 will be displayed.   
Pressing the print key will exit back to the menu T3.

       

T4  

A to D

 Displays Raw counts where a 1mV/V signal from the scale

will display 25,000 counts.

When in angle mode Down GRS/NET will cycle Pitch, Roll 
and Raw counts. 

 

T5

 Serial

Serial communications can be verified by connecting pins
2 and 3 on the serial port. A single character will be echoed 

 and pass or fail will be displayed.    

       

T6

 Setup data

Setup data will be sent out on the printer port. 

       

T7

 Default

Resets the meter back to factory defaults clearing all  

calibration and setup data. "r you sure?" will be displayed 

then press the Tare key to exit without Defaulting, or press 

the Print key to reset the meter to factory default.  

      

END

Exits back to the main Menu.

5

.0 TESTING PROCEDURES

CAMBRIDGE SCALE WORKS, INC.

1

5

Manual P/N 5999-1011-02

Содержание CSW-10AT-B

Страница 1: ......

Страница 2: ... 2 3 Scale Menu Definitions 8 9 2 4 Numeric Entries 10 2 5 Set Up Parameters 10 3 0 Calibration Procedures 11 3 1 Calibration Menu Definitions 11 3 2 Calibration 12 3 3 Linearity Correction 12 4 0 Communications Setup 13 4 1 Communications Menu Definitions 13 14 5 0 Testing Procedure 15 5 1 Testing Menu Definitions 15 6 0 Warranty 16 7 0 Assistance 17 CAMBRIDGE SCALE WORKS INC 2 Manual P N 5999 10...

Страница 3: ... in the battery holder with the contacts pointing down then slide down to lock into position Meter Connections Communications port Signal connection PIN DESCRIPTION PIN DESCRIPTION 3 TXD 1 EXC 4 5V 2 SIG 5 SGND 5 SHIELD LOAD CELL 6 EXC 7 SIG CAMBRIDGE SCALE WORKS INC 3 Manual P N 5999 1011 02 ON OFF SWITCH CHARGER POWER Charge indicating LED ...

Страница 4: ...s the Gross weight value into the Tare display and switches to the Net display mode If the Tare key is pressed and held for 5 seconds the current Tare value will be displayed lb kg Toggles the display between pounds and kilograms PRINT Outputs the displayed weight data to the RS 232 Port Note All keys are disabled when the scale is in motion or overload 1 2 Error Messages ScnEg When the weight is ...

Страница 5: ...f the power adapter to a 110VAC outlet Next connect the other end of the adapter to the power jack on the rear of the meter The charge indicating led will be orange while the battery is being charged and turned off when the battery is fully charged Turn the on off switch to the ON postion and the meter will be fully operational as the battery is being charged CAMBRIDGE SCALE WORKS INC 5 Manual P N...

Страница 6: ... ZERO SHOW RAW COUNTS THEN 0 SHOW RAW COUNTS THEN 0 SHOW ACTUAL WEIGHT C3 SPAN POINT C4 SPAN POINT C5 SPAN POINT C6 SPAN POINT C7 SPAN POINT END R1 BAUD RATE R2 Output Format R3 Output Type R4 Receiver mode 1200 8 N 1 19200 8 N 1 WT ONLY G N GTN Key Polled Continuous Disabled Rec Remote END T1 VERSION T2 DISPLAY LIGHT SEGMENTS SOFTWARE VERSION T3 BUTTONS b1 b4 b5 EXITS T4 A TO D A D RAW COUNTS T5 ...

Страница 7: ...ATE LEFT lb kg KEY IS USED TO NAVIGATE RIGHT PRINT KEY IS USED TO ENTER DATA AND RETURNS 2 1 SOFTWARE NAVIGATION FLOWCHART CONTINUED S9 Motion Display Normal Blank Dashes Normal S10 ZERO BAND 1 99 DIVISIONS 2 S11 ZERO DELAY 0 99 UPDATES 2 1 99 DIVISIONS 1 1 99 1 4secinc 0 lb kg lb kg lb kg S12 Zero Tracking S13 Tracking DELAY S14 Display lb kg S15 SCALE ID S16 BRIGHTNESS S17 SLEEP MODE END 1 99 1 ...

Страница 8: ...r 1 9 of scale capacity S6 Filter 0 7 3 0 is the fastest response or least filtering and 7 is the slowest response or most filtering S7 Motion Band 1 to 99 updates 2 the weight display must be stable within the selected number of updates for the motion indicator to be turned off S8 Motion Delay 0 to 99 updates 4 the weight display must be within the motion band for the selected number of updates f...

Страница 9: ... before it will be automatically Zeroed S14 Lb Kg Lb KG Allows the indicator to be switched between pounds and kilograms by pressing the lb kg key LB This sets the display to Pounds only Kg This sets the display to Kilograms only S15 Scale ID 1 to 99 scale ID used in RF link output S16 Brightness 0 to 15 adjusts the LED display intensity 15 is the brightest S17 Sleep Mode 0 to 30 minutes 20 The di...

Страница 10: ...anges are made to the Scale or Calibration Parameters With the display indicating ScAlE press the GRS NET key to move down This allows the operator to change any of the scale Parameters S1 thru S17 For example press the GRS NET key to move down until S2 is displayed S2 is used to set the Capacity of the Scale Press the lb kg key to move right in the S2 Parameter The current capacity will be displa...

Страница 11: ...own test weight press the ZERO key to enter the correct weight Use the steps described in section 2 4 for numeric entry When the weight is correct press the Print key to enter the new value C4 to C7 Span points C4 to C7 are for linearity correction they can be used in order and in any quantity or not at all if no correction is necessary You may also return to C4 to C7 later and add a new correctio...

Страница 12: ... move left the next digit will flash Repeat the steps until the correct weight is entered Press the PRINT key to record the data The display will return to C3 Press the GRS NET key until End is displayed then press PRINT Calibration is now complete 3 3 Linearity Correction If Linearity Correction is needed Press the GRS NET key from the calibration menu to move down to C4 Press the lb kg key to mo...

Страница 13: ... GR Sp Six ASCII Characters indicated weight Sp lb or kg CR LF 14 bytes total output R3 Output Type 0 Output on comand standard print Output as selected by R2 output format If Q is received on the serial port the scale will output the same a if the PRINT key were pressed the same holds true for Z Zero U lb kg D GRS NET T Tare 1 Slave display output numeric only continuous Stx Six ASCII Characters ...

Страница 14: ...ys are disabled except the PRINT key 2 Remote control Receives R3 3 RF link output string displaying data as it appears on the scale display and allows full control of all scale meter functions End Exits back to main Menu indicates Factory Default settings 4 1 Communications Menu Definitions Continued CAMBRIDGE SCALE WORKS INC 14 MANUAL P N 5999 1011 02 ...

Страница 15: ...counts When in angle mode Down GRS NET will cycle Pitch Roll and Raw counts T5 Serial Serial communications can be verified by connecting pins 2 and 3 on the serial port A single character will be echoed and pass or fail will be displayed T6 Setup data Setup data will be sent out on the printer port T7 Default Resets the meter back to factory defaults clearing all calibration and setup data r you ...

Страница 16: ...ts products beyond that stated above Warranty replacement parts and or repair services are performed at the factory in Cumberland Maryland or by an authorized Service Group approved by CAMBRIDGE Warranty does not include travel expense if a factory technician is requested to perform repairs at a location other than the factory It s the user s responsibility to follow the proper set up calibration ...

Страница 17: ...istance with your Model CSW 10AT B Meter please contact us at CAMBRIDGE SCALE WORKS INC 115 West Mary Street Cumberland MD 21502 Phone 301 724 4082 Fax 301 724 4964 7 0 Assistance CAMBRIDGE SCALE WORKS INC 17 Manual P N 5999 1011 02 ...

Отзывы:

Похожие инструкции для CSW-10AT-B