CAMAG UV LAMP 4 Скачать руководство пользователя страница 1

INSTRUCTION MANUAL

UV LAMP 4

UV CABINET 4

Содержание UV LAMP 4

Страница 1: ...INSTRUCTION MANUAL UV LAMP 4 UV CABINET 4...

Страница 2: ...n 5 2 1 UV Lamp 4 5 2 2 UV Cabinet for UV Lamp 4 5 3 Getting started 7 3 1 Normal Operation 7 3 2 Safety Mode 7 3 3 Error Mode 7 4 Service and maintenance 8 4 1 Replacing fluorescent tubes 8 4 2 Clean...

Страница 3: ...detection method UV intensity and complete elimination of visible light are less critical CAMAG Wavelength statement In literature both UV 365 nm and UV 366 nm are used to describe the UV light spect...

Страница 4: ...ure range outlined in the chapter Technical Data Before installation and use the instrument should be acclimated properly Use a damp lint free cloth for cleaning the instrument surface Do not employ a...

Страница 5: ...rective 2002 96 EC To properly recycle the instrument or its parts you are requested to send the equipment back to the distributor producer or an adequate collection system at the end of its life This...

Страница 6: ...seconds Button LED is blinking 2 2 UV Cabinet for UV Lamp 4 The UV Cabinet is used for inspecting TLC plates under UV light in a bright environment A glass filter protects the eyes against UV light re...

Страница 7: ...Instruction Manual November 2017 6 Figure 1 Place the UV lamp in the recess provided on top of the UV Cabinet 4 Figure 2...

Страница 8: ...e 4 3 2 Safety Mode The Lamp has a built in safety mode which is active when either the lamp is tilted more than 30 or when the lamp temperature reaches a certain limit It is indicated by the blinking...

Страница 9: ...g it approx 90 Insert the pins of the new tube into the sockets and turn the tube about 90 again Tighten up the screws in the filter holding plate Reconnect the UV lamp to the power cable and check th...

Страница 10: ...Instruction Manual November 2017 9 Figure 3 Attention Your eyes are protected from reflected UV light with this filter Never use the UV Cabinet 4 unless the filter is properly installed...

Страница 11: ...ideally adjusted for the use of TLC HPTLC plates Please be aware that direct and reflected UV light can be harmful and has to be avoided Average intensities of direct and indirect UV light can be foun...

Страница 12: ...icator The intensity of the reflection for 20 x 10 cm plates and plates with UV indicator is slightly lower 5 3 UV Cabinet for UV Lamp 4 Dimensions l x w x h 420 x 296 x 280 mm Weight 4 0kg CAMAG Swit...

Страница 13: ......

Отзывы: