800-572-1028 caltestelectronics.com 7
WARNING
At the time of powering on the probe, the input leads must not be
connected to an item to be tested. Never operate the probe with the
case open.
Inspection Procedure
1.
Connect the BNC output connector to the vertical input of the
oscilloscope.
2.
Power on the probe.
3.
Set the attenuation setting on the oscilloscope to match the
probe (10x).
4.
Set the oscilloscope input to DC coupling and .5V/div. Center
the trace on the display.
5.
Connect the input pins of probe to a function generator and
select a square-wave output of 10 V amplitude and 100 kHz
frequency.
6.
A 100 kHz square-wave of 1V amplitude will be displayed on the
screen of the oscilloscope. This indicates the probe is working
properly.
Getting Started
1.
Attach the desired SMD accessories to the inputs
2.
Connect the BNC output connector to the vertical input of the
oscilloscope.
3.
Power on the probe.
4.
Set the attenuation setting on the oscilloscope to match the
probe (10x).
5.
Connect the inputs to the circuit to be tested.