Twister II - Maintenance 109
P/N 113564 Rev. 2
Twister II Robot User’s Manual
Caliper Life Sciences, Inc.
Changing the Assay Engine Twister II Fuses
3
Remove the two 2.5Amp fuses.
shows the fuse
housing for the Assay Engine Twister II.
Figure 88. Assay Engine Twister II Fuses
4
Replace both fuses with the same part number fuses.
5
Confirm that the fuses are seated securely in the fuse housing.
6
Snap the fuse housing back into the power entry port.
Moving the Twister II
If you need to relocate the Twister II:
1
Remove all racks from the pods and materials from the racks.
2
Turn OFF the power.
3
Unplug the power cord from the power outlet and from the
receptacle on the side of the pod.
4
Disconnect the communications cable from the side of the pod.
5
Prepare a space to accommodate the Twister II during the move
(an empty area on the lab bench or a sturdy cart).
6
Disconnect all the pods from the robot canister. Reverse the
hardware installation steps in the Robot Installation Instructions.
WARNING
The Twister II weighs approximately 50 pounds (22.7 kg) and
should be lifted by at least two people. Take the proper precautions
to avoid injury.