Distance and Spot Size

Ensure the sensor is positioned
so that it can only detect infrared
radiation from the target.

The size of the area (spot size) to
be measured determines the
distance between the sensor and
the target. The spot size must
not be larger than the target. The
sensor should be mounted so
that the measured spot size is
smaller than the target.

We normally recommend the
target is at least twice the size of
the given measured spot for
maximum accuracy.

PREPARATION

Ambient Temperature

The sensing head is designed to operate in ambient temperatures from 0°C to 180°C.
The main enclosure is designed to operate in ambient temperatures from 0°C to 70°C.
Avoid thermal shock. Allow 20 minutes for the unit to adjust to large changes in ambient
temperature.

Atmospheric Quality

Smoke, fumes or dust can contaminate the lens and cause errors in temperature
measurement. In these types of environment the air purge collar should be used to help keep
the lens clean.

Electrical Interference

To minimise electromagnetic interference or ‘noise’, the sensor should be mounted away from
motors, generators and such like.

Wiring

Check the distance between the sensor and the indicating/controlling device. If necessary, the
sensor can be ordered with a longer cable attached. Contact Calex for information about
extending the cable.

Power Supply

A suitable intrinsically safe isolator must be used. See

Specifications

for the supply voltage,

current and safety requirements.

MECHANICAL INSTALLATION

Mounting

The sensing head is supplied with 2 mounting nuts. The sensor can be mounted on brackets
or cut outs of your own design, or you can use the fixed and adjustable mounting bracket
accessories, which are shown in the following diagram.

Sensor

CORRECT

Target larger

than spot size

Target smaller

than spot size

INCORRECT

Background

8

Содержание ExTempMini Series

Страница 1: ...ExTempMini Series Intrinsically Safe Miniature Infrared Temperature Sensor with Separate Electronics Module Operator s Guide...

Страница 2: ...ty i IMPORTANT INFORMATION FOR USE Voltage MUST be supplied by a suitably rated safety isolator For re configuration of the sensor the LCT MUST be connected in the safe area behind the protection of a...

Страница 3: ...rsist for a short period NO Group Definition Supported Gas IIA e g Propane IIB e g Ethylene IIC e g Hydrogen Dust IIIA Combustible flyings NO IIIB Non conductive dusts NO IIIC Conductive dusts NO Tabl...

Страница 4: ...ge may be rescaled between limits 20 C and 1000 C via optional adapters and software Sensing Head Cable Length m e g 5 5 m 10 10 m 25 25 m maximum length The sensing head and electronics module are ca...

Страница 5: ...ntal Rating IP65 NEMA 4 IP65 NEMA 4 Ambient Operating Temperature Range 0 C to 70 C 0 C to 180 C Ambient Pressure Range 80 kPa 0 8 bar to 110 kPa 1 1 bar 80 kPa 0 8 bar to 110 kPa 1 1 bar Relative Hum...

Страница 6: ...8 V Maximum Input Current Ii 93 mA Maximum Input Power Pi 0 651 W Maximum Internal Capacitance Ci 5 17 nF Maximum Internal Inductance Li 1 99 H ATEX Certificate Number CML 22ATEX2007X IECEx Certificat...

Страница 7: ...ended sensing head cable 25 m max factory fitted contact Calex for information about extending the cable Output cable 25 m max factory fitted cable alternatively customer may supply and fit output cab...

Страница 8: ...ontaminate the lens and cause errors in temperature measurement In these types of environment the air purge collar should be used to help keep the lens clean Electrical Interference To minimise electr...

Страница 9: ...way from the lens It must be screwed on fully There are two models of air purge collar to suit the different optics see below Air flows into the hose barb fitting and out of the front aperture Air flo...

Страница 10: ...adapter PC with configuration software Load measurement instrumentation 220 Vs Do not open PWR PWR Sensing head Ta 20 C to 180 C Certified Intrinsically Safe isolator e g connections shown are for MT...

Страница 11: ...MODBUS The optional LCT 485 Network Interface not shown allows an ExTemp sensor to be connected to an RS 485 Modbus RTU network for temperature measurement configuration and data acquisition Each LCT...

Страница 12: ...ing can be applied by setting Hold Mode to Peak or Valley and setting the hold period This is useful if the temperature reading is interrupted by gaps between moving objects or by an obstruction Defau...

Страница 13: ...for further advice LENS CLEANING Keep the lens clean and dry at all times Any foreign matter on the lens would affect measurement accuracy Blow off loose particles if not using the air purge accessory...

Страница 14: ...safe side of a suitable certified Intrinsically Safe isolator Do not attempt to repair a faulty unit Contact the vendor to arrange a return ELECTRICAL CONNECTION Connect the device to the 4 to 20 mA...

Страница 15: ...ARE See Configurable Parameters and the Configuration Software diagram opposite If the Settings icon is greyed out the software is locked Unlock the software to allow access to the Settings menu The d...

Страница 16: ...twork interface Select COM port then select LCT 485 on the next screen Emissivity Compensation Temperature List Temperature Display Output Range Output Processing Default Password 1234 Connection List...

Отзывы: