Caldera Benesta CAL-TR1511 Скачать руководство пользователя страница 6

 

 

Interconnections conditions require the applied parts of other medical 

electrical equipment used within the configuration for hysteroscopic 

application to be type BF applied parts or type CF applied parts 

 

This equipment is designed and tested to minimize interference with 

other electrical equipment. However, if interference occurs with other 

equipment it may be corrected by one or more of the following 

measures: 

1.

 

Reorient or relocate this equipment, the other equipment, or both. 

2.

 

Increase the separation between the pieces of equipment. 

3.

 

Connect the pieces of equipment into different outlets or circuits. 

4.

 

Consult a biomedical engineer. Electrical safety testing should be 

performed by a trained biomedical engineer or other qualified 

person. 

 

It is the responsibility of the person(s) installing, utilizing and 

decommissioning the Benesta

TM

 Tissue Removal Device to ensure that 

the electromagnetic environment does not exceed or deviate from the 

specifications listed in Tables 1-4.  

 

The following tables provide information on the electromagnetic 

environment that the Benesta

TM

 Tissue Removal Device is capable of 

operating in safely. 

Table 1. Declaration of Electromagnetic Emissions 

The Benesta

TM 

Tissue Removal Device is intended for use in the electromagnetic environment 

specified below. The customer or the user of the Benesta

TM

 Tissue Removal Device should assure that 

it is used in such an environment. 

Emissions test 

Test Level/ 

Limits 

Electromagnetic Environment / Recommendations 

RF emissions CISPR 11  Class A 

The EMISSIONS characteristics of this equipment make 

it suitable for use in industrial areas and hospitals 

(CISPR 11 class A). If it is used in a residential 

environment (for which CISPR 11 class B is normally 

required) this equipment might not offer adequate 

protection to radio-frequency communication services. 

The user might need to take mitigation measures, such 

as relocating or re-orienting the equipment 

Harmonic  distortion 
IEC 61000-3-2 
IEC 61000-3-2 

Class A 

Not Applicable - device is battery operated 

Voltage fluctuations/ 

flicker emissions 
IEC 61000-3-3 

 Pst  ≤1.0 
 Plt  ≤ 0.65 
 dc  ≤ 3.3% 
 dmax ≤4.0% 

Not Applicable - device is battery operated 

 

Table 2. Declaration of Electromagnetic Immunity 

Immunity Test 

Test Level / Limits 

Electromagnetic Environment 

/ Recommendations 

Electrostatic 

Discharge (ESD) 

IEC 61000-4-2 

8 kV (±) Contact 

15 kV (±) Air  

Floors should be wood, 

concrete, or ceramic tile. If floors 

are covered with synthetic 

material, the relative humidity 

should be at least 30% 

Electrical Fast 

Transients 

IEC 61000-4-4 

±2 kV AC/DC power supply 

±1 kV input/output lines 

Not Applicable – device is 

battery-operated 

Surge 

IEC 61000-4-5 

±1 kV to line(s) 

±2 kV to line(s) to earth  

Not Applicable – device is 

battery-operated 

Radio Frequency 

(RF) Common Mode 

IEC 61000-4-6 

3 Vms  

150kHz to 80MHz  

Not Applicable – device is 

battery-operated 

Power Frequency 

Magnetic Field 

IEC 61000-4-8 

30 A/m  

Power frequency magnetic fields 

should be at levels typical 

commercial or hospital 

environments.  

Voltage Dips and 

Interruptions 

IEC 61000-4-11 

<5% U

T 

(>95% dip in U

T

) for 0.5 cycle 

40% U

T 

(60% dip in U

T

) for 5 cycles 

70% U

T 

(30% dip in U

T

) for 25 cycles 

<5% U

T 

(>95% dip in U

T

) for 5 sec  

Not Applicable – device is 

battery-operated 

U

T

 is the A.C. mains voltage prior to application test level 

 

Table 3. Guidance and Manufacturer’s Declaration of Electromagnetic 

Immunity 

The Benesta

TM 

Tissue Removal Device is intended for use in the electromagnetic environment 

specified below. The customer or the user of the Benesta

TM 

Tissue Removal Device should assure that 

it is used in such an environment. 

Immunity test 

IEC 60601 

test level 

Compliance 

level 

Electromagnetic environment – guidance 

Conducted 

Radiated 

RF 

IEC 

61000-4-3 

3 Vrms 

3 V/m 

80 MHz to 

2.5 GHz 

3 Vrms 

3 V/m 

Portable and mobile RF communications equipment 

should be used no closer to any part of the Benesta

TM 

Tissue Removal Device, including cables, than the 

recommended separation distance calculated from the 

equation applicable to the frequency of the transmitter. 

 

Recommended separation distance (in meters) 

= 1.2√80 MHz to 800 MHz 

= 2.3√800 MHz to 2.5 GHz 

where is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and d is the recommended separation 

distance in meters (m).  

 

Field strengths from fixed RF transmitters, as 

determined by an electromagnetic site survey, should be 

less than the compliance level in each frequency range. 

 

Interference may occur in the vicinity of equipment 

marked with the following symbol: 

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people. 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones 

and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be 

predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 

transmitters, an electromagnetic site survey should be considered. If the measured field strength in the 

location in which the Benesta

TM

 Tissue Removal Device is used exceeds the applicable RF compliance 

level above, the Benesta

TM

 Tissue Removal Device should be observed to verify normal operation. If 

abnormal performance is observed, additional measures may be necessary, such as re-orienting or 

relocating the Benesta

TM

 Tissue Removal Device. 

 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 

 

Содержание Benesta CAL-TR1511

Страница 1: ...ns for Use CAL TR1511 1 pack CAL TR1513 3 pack Manufactured by Caldera Medical Inc 5171 Clareton Drive Agoura Hills CA 91301 U S Toll Free 866 4 CALDERA Telephone 818 879 6555 Fax 818 879 6556 www cal...

Страница 2: ...Not Re use Indicates a medical device that is intended for one use or for use on a single patient during a single procedure 1051 5 4 2 Consult Instructions for Use Indicates the need for the user to...

Страница 3: ...presence of cancer cells Exercise extreme caution when resecting tissue in patients who have implants that extend into the uterine cavity Do not use the BenestaTM Tissue Removal Device to resect tissu...

Страница 4: ...e IV Pole pressure cuff or fluid management system FMS Distending media Vacuum source suction canister and tissue trap BenestaTM Hysteroscope and outflow channel Inflow tubing Y tubing Hysteroscope an...

Страница 5: ...ng hard into target pathology Excessive leverage on the device does not improve cutting performance and may result in decreased performance and or unintended excision of tissue pathology 13 Excised ti...

Страница 6: ...to earth Not Applicable device is battery operated Radio Frequency RF Common Mode IEC 61000 4 6 3 Vms 150kHz to 80MHz Not Applicable device is battery operated Power Frequency Magnetic Field IEC 61000...

Страница 7: ...by Caldera Medical at its option of any product found to be defective during the warranty period Damage inflicted to a product by the user will result in additional charges and may void the warranty T...

Отзывы: